10.4.3 Calculate Limit
Quartile
When the click rate N is determined it is possible to calculate the click limit
Lq by increasing the relevant limit (quasi-peak) L for continuous
disturbances with:
●
44 dB for N<0,2, or
●
20 log (30/N) dB for 0,2<N<30
Therefore
N
Log
L
L
q
30
20
⋅
+
=
10.4.4 Measurement vs.
Lq limit
Now the test can be done at the 4 fixed frequencies dictated by the
standard: 150 kHz, 500 kHz, 1.4 MHz, 30 MHz, and the results compared
with the limit Lq.
If less than 25% of the clicks measured exceeds the limit Lq, then the test
is positive, otherwise it fails.
The PMM 9010 can now produce an extensive report with all the relevant
data.
10.5 Start
As soon as the Setup has been done (see § 9.5), it is possible to start the
test.
First of all the PMM 9010 will determine the click rate N, and – if all the
conditions apply – after that the analyzer will immediately continue
measuring the number of clicks exceeding the upper quartile limit L
q
.
Simply select
Start
: the PMM 9010 will automatically take care of all the
necessary steps, reporting at the end of the test all the relevant results.
During the test, the evolution of the measurements can be continuously
controlled in the Data and Details windows, where all the info are duly
provided.
At any time, it is possible to
Stop
the measure or to
Pause
it.
The PMM 9010 does the determination of the click rate at the two
frequencies of 150 kHz (
Step 1
) and 500 kHz (
Step 3
).
Summary of Contents for PMM 9010 Series
Page 24: ...XXIV Safety considerations This page has been left blank intentionally ...
Page 42: ...2 10 Installation This page has been left blank intentionally ...
Page 66: ...6 6 Manual Mode Operating Instructions This page has been left blank intentionally ...
Page 72: ...7 6 Applications This page has been left blank intentionally ...
Page 92: ...9 14 General Information This page has been left blank intentionally ...
Page 112: ...10 20 Click Mode Operating Instructions This page has been left blank intentionally ...
Page 145: ...PMM 9030 9060 9180 Frequency Extension 11 33 Screwing the N N Adapter to BL 01 connector ...
Page 170: ...11 58 PMM 9030 9060 9180 Frequency Extension This page has been left blank intentionally ...
Page 178: ...12 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally ...
Page 194: ...13 16 APD Mode Operating Instructions This page has been intentionally left blank ...
Page 230: ...14 36 Remote control This page has been intentionally left blank ...
Page 284: ...Suggerimenti Commenti Note Suggestions Comments Note ...