APD Mode Operating Instructions
13-5
13.3.4 Limit 2 P
Moving the highlighter cursor on the fourth line of the Setup page the
threshold percentage of probability, expressed in power of ten, for the
second limit can be entered using the function keys on the right.
The range for this limit is between 10
0
and 10
-8
dBµV (default: 10E-4).
Only the positive number of the exponent needs to be entered.
To set 10
0
a value of 0.1 must be digited.
Enter a 0 (zero) to disable the limit.
If the limit is disabled a series of dashes ( ---- ) is displayed.
13.3.5 YY Offset
When the YY Offset function is activated, only the frequencies with a level
over the higher between the two limits (+ offset) will be investigated during
the test.
Moving the highlighter cursor on the fifth line of the Setup page the offset,
expressed in dB, related to the first Limit 1 E can be entered using the
function keys on the right.
The range for this value is between 0 dB and 150 dB.
To set 0 dB
a value of 0.01 must be digited.
Enter a 0 (zero) to disable this function.
If the limit is disabled a series of dashes ( ---- ) is displayed (default).
13.3.6 Start Frequency
Moving the highlighter cursor on the eighth line of the Setup page the
starting frequency for the searching sweep, expressed in MHz, can be
entered using the function keys on the right.
The range for this value is between 30 and 3000 MHz.
When a value outside the range is entered the default 1000 MHz is
automatically set.
13.3.7 Stop Frequency
Moving the highlighter cursor on the ninth line of the Setup page the end
frequency for the searching sweep, expressed in MHz, can be entered using
the function keys on the right.
The range for this value is between 30 and 3000 MHz.
When a value outside the range is entered the default 3000 MHz is
automatically set.
13.3.8 Num. of freqs
Moving the highlighter cursor on the tenth line of the Setup page the number
of frequencies for the searching sweep, the ones that will be further
investigated, can be entered using the function keys on the right.
The range for this value is between 2 and 50.
When a value outside the range is entered the default number of 5 is
automatically set.
Summary of Contents for PMM 9010 Series
Page 24: ...XXIV Safety considerations This page has been left blank intentionally ...
Page 42: ...2 10 Installation This page has been left blank intentionally ...
Page 66: ...6 6 Manual Mode Operating Instructions This page has been left blank intentionally ...
Page 72: ...7 6 Applications This page has been left blank intentionally ...
Page 92: ...9 14 General Information This page has been left blank intentionally ...
Page 112: ...10 20 Click Mode Operating Instructions This page has been left blank intentionally ...
Page 145: ...PMM 9030 9060 9180 Frequency Extension 11 33 Screwing the N N Adapter to BL 01 connector ...
Page 170: ...11 58 PMM 9030 9060 9180 Frequency Extension This page has been left blank intentionally ...
Page 178: ...12 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally ...
Page 194: ...13 16 APD Mode Operating Instructions This page has been intentionally left blank ...
Page 230: ...14 36 Remote control This page has been intentionally left blank ...
Page 284: ...Suggerimenti Commenti Note Suggestions Comments Note ...