Click Mode Operating Instructions
10-9
10.5.1 Stop and Pause
If Stop has been selected during a measurement, the test can then be
aborted or it is possible to go immediately to the next step.
If Paused, the test can be resumed at any convenient time.
10.6 Report
After a Click test the PMM 9010 reports all the relevant data.
Due to the complexity of the test and to the various conditions that may
happen during the measurement, several reports may be issued.
Here following some examples well representing the completeness and the
accuracy of the analyzer.
10.6.1 Fail during
determination of
the click rate N
In case a fail is recorded during the determination of the click rate N at the
frequency of 150 kHz, a report like the following is displayed (which is just
an example with a not-realistic N number):
If the fail occurs during the determination of the click rate at 500 kHz, the
report may be like this:
10.6.2 Report after a
successful test with
less than 5
instantaneous
switching at one
frequency
If the determination of the click rate N shows that at a given frequency (500
kHz in this examples) there are less than 5 instantaneous switching
operations, there is no need of any further test at that frequency or
frequencies (the investigation frequencies are 150 kHz for the interval from
150 to 500 kHz, and 500 kHz for the interval from 500 kHz to 30 MHz).
In the example, L
q
is calculated for 150 kHz and no tests are done at the
other frequencies, as the number in instantaneous switching at 500 kHz
during the observation time is less than 5.
To complete the test of the example above, only the steps 1, 2 and 3 were
performed.
Summary of Contents for PMM 9010 Series
Page 24: ...XXIV Safety considerations This page has been left blank intentionally ...
Page 42: ...2 10 Installation This page has been left blank intentionally ...
Page 66: ...6 6 Manual Mode Operating Instructions This page has been left blank intentionally ...
Page 72: ...7 6 Applications This page has been left blank intentionally ...
Page 92: ...9 14 General Information This page has been left blank intentionally ...
Page 112: ...10 20 Click Mode Operating Instructions This page has been left blank intentionally ...
Page 145: ...PMM 9030 9060 9180 Frequency Extension 11 33 Screwing the N N Adapter to BL 01 connector ...
Page 170: ...11 58 PMM 9030 9060 9180 Frequency Extension This page has been left blank intentionally ...
Page 178: ...12 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally ...
Page 194: ...13 16 APD Mode Operating Instructions This page has been intentionally left blank ...
Page 230: ...14 36 Remote control This page has been intentionally left blank ...
Page 284: ...Suggerimenti Commenti Note Suggestions Comments Note ...