Setup and Panel Instructions
3-3
3.4 Unit
Entering in the
Unit
menu it is possible to change the indicated unit used to
display the measured levels.
The available units are
dBµV
and
dBm
.
In a 50 Ohm system the relationship between dBm and dB
μ
V is:
dB
μ
V = dBm + 107
i.e., 0 dBm (1 mW) = 107 dB
μ
V
3.5 RF OUT
The
tracking generator
is an internal, high stability and accuracy, 50 Ohm
RF generator ranging from 10 Hz to 50 MHz.
Activating the
Tracking On
function the generator is always tuned at the
same PMM 9010 measurement frequency and scans the range together
with the receiver. This is the standard way a tracking generator works in all
spectrum analyzers.
If the Tracking On function is disabled, the generator becomes a CW signal
source tuned at the frequency set under the
RF OUT Freq
window.
As usual, to set a given frequency it is therefore enough to edit the value
with the soft keys and the left and right arrows, confirming the selections
entering the units value (kHz, MHz or GHz).
Pressing 0 as the first figure the decimal dot appears automatically.
In both operating modes (tracking or fixed frequency) the output level can
be set between 60,0 and 90,0 dBµV with 0,1dB steps using the
RF OUT
Level
button.
If a higher or lower level is needed, the User shall adopt either an external
amplifier or an external attenuator.
The tracking generator is extremely useful for several applications: first of
all it is essential to calibrate the receiver itself thanks to the automatic
internal routines specifically developed to this extent, then it can also be
used to transform the PMM 9010 in a scalar network analyzer, helping a lot
in designing and testing RF filters, active stages and a lot of other circuits.
3.6 Panel
The Panel function allows the User to store up to 2 different setups that
can be recalled any time.
Pressing
Save #1
or
Save #2
the actual setup is stored in the internal
memory; with the two
Recall
buttons the corresponding setup is loaded to
the receiver.
Please remember each time one of the Save buttons is pressed, the stored
set-up is overwritten by the new one: the saved setups are therefore kept
memorized in the receiver until a new set-up is stored in the same
memory.
The
Default
button can be used to load a default standard setup saved
into the memory at the factory.
Summary of Contents for PMM 9010 Series
Page 24: ...XXIV Safety considerations This page has been left blank intentionally ...
Page 42: ...2 10 Installation This page has been left blank intentionally ...
Page 66: ...6 6 Manual Mode Operating Instructions This page has been left blank intentionally ...
Page 72: ...7 6 Applications This page has been left blank intentionally ...
Page 92: ...9 14 General Information This page has been left blank intentionally ...
Page 112: ...10 20 Click Mode Operating Instructions This page has been left blank intentionally ...
Page 145: ...PMM 9030 9060 9180 Frequency Extension 11 33 Screwing the N N Adapter to BL 01 connector ...
Page 170: ...11 58 PMM 9030 9060 9180 Frequency Extension This page has been left blank intentionally ...
Page 178: ...12 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally ...
Page 194: ...13 16 APD Mode Operating Instructions This page has been intentionally left blank ...
Page 230: ...14 36 Remote control This page has been intentionally left blank ...
Page 284: ...Suggerimenti Commenti Note Suggestions Comments Note ...