Remote control
14-11
14.6.2 SETTING
Commands
Behaviour
ASBK
This command aborts a sweep currently in progress previously started by command
SSDO
or
SSFD
. The reply is either:
•
SBK=OK
which means the sweep has been stopped and all setting
commands are available again. In case the sweep was triggered by
SSDO
,
with partial data, data are available for uploading by the command “
?FSF 0
”
•
SBK=SERR
which informs that the sweep was not in need to be stopped
ASPA
This command suspends a sweep currently in progress previously started by
command
SSDO
or
SSFD
. There is no reply unless a sweep is not in progress in
which case the reply is:
•
SPA=SERR
which informs that the sweep was not in need to be paused
ASRE
This command resumes a sweep previously paused by command
ASPA
. There is no
reply unless a sweep is not in progress in which case the reply is:
•
SRE=SERR
which informs that the sweep was not in need to be resumed
S3PRC
This setting command
#S3PRC*
sets the PMM 9010/03P/30P/60P to conducted
mode (BANDs A & B). The reply is:
•
3PR=OK
(The PMM 9010/03P/30P/60P has been set for conducted mode )
•
3PR =BERR
(The command has been ignored for one of the following:
•
The connected apparatus is not a PMM 9010/03P/30P/60P
•
The instrument is not idling and therefore cannot switch to a different
mode. In such an event, stop the current process (Manual, Analyzer or
Sweep) and put it back to Idle.
•
3PR =SERR
(The command contains a syntax error )
S3PRR
This setting command
#S3PRR*
sets the PMM 9010/03P/30P/60P to radiated mode
(BANDS C for 9010/03P and BANDS C, D & E for 9010/30P). The reply is:
•
3PR=OK
(The PMM 9010/03P/30P/60P has been set for radiated mode )
•
3PR =BERR
(The command has been ignored for one of the following:
•
The connected apparatus is not a PMM 9010/03P/30P/60P
•
The instrument is not idling and therefore cannot switch to a different
mode. In such an event, stop the current process (Manual, Analyzer or
Sweep) and put it back to Idle.
•
3PR =SERR
(The command contains a syntax error )
SAAT a
This setting command sets Attenuator, for Analyzer Mode, to the value indicated by
(a) which can be 0 to 35 (55). Automatic Attenuator feature is disabled. If (a) is a
negative figure then the Automatic Attenuator feature is turned on. The reply is
AAT=OK
which acknowledges the command has been granted or
AAT=SERR
if the
command has been ignored.
Example:
# AAT 15*
(set attenuator to 15dB and disable Automatic Attenuator
feature)
Example:
# AAT -1*
(Enable Automatic Attenuator feature)
SADT b
This setting command sets the detector for Analyzer Mode. The argument (b) should
be an index representing the detector as follows:
o
1 Peak
o
2 Avg
o
3 Rms
The reply is
SADT =OK
which acknowledges the command has been granted or
SADT =SERR
if the command has been ignored.
Example:
# SADT 1*
sets the detector to Peak for Analyzer Mode
Summary of Contents for PMM 9010 Series
Page 24: ...XXIV Safety considerations This page has been left blank intentionally ...
Page 42: ...2 10 Installation This page has been left blank intentionally ...
Page 66: ...6 6 Manual Mode Operating Instructions This page has been left blank intentionally ...
Page 72: ...7 6 Applications This page has been left blank intentionally ...
Page 92: ...9 14 General Information This page has been left blank intentionally ...
Page 112: ...10 20 Click Mode Operating Instructions This page has been left blank intentionally ...
Page 145: ...PMM 9030 9060 9180 Frequency Extension 11 33 Screwing the N N Adapter to BL 01 connector ...
Page 170: ...11 58 PMM 9030 9060 9180 Frequency Extension This page has been left blank intentionally ...
Page 178: ...12 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally ...
Page 194: ...13 16 APD Mode Operating Instructions This page has been intentionally left blank ...
Page 230: ...14 36 Remote control This page has been intentionally left blank ...
Page 284: ...Suggerimenti Commenti Note Suggestions Comments Note ...