14-34
Remote control
14.7.1.4
Analyzer MAX-
HOLD related
COMMANDs
Query
Command Description
?MHF
Max
Hold
Function
This query command
#?MHF *
sends back a string expressing the Analyzer Max-Hold
function STATUS.
The reply is either:
•
MHF=OFF.
Analyzer Max-Hold function is disabled and the sent spectrum reflects
the current one.
•
MHF=ON .
Analyzer Max-Hold function is enabled and the sent spectrum reflects
the maximum value recorded, frequency by frequency, since it was last cleared.
?MHP
Max
Hold
Pause
This query command
#?MHP *
sends back a string expressing the Analyzer Max-Hold
PAUSE Status.
The reply is either:
•
MHP=OFF.
Analyzer Max-Hold function is active and every new spectrum
updates the maximum value.
•
MHP=ON .
Analyzer Max-Hold function is paused (suspended). All incoming
values will not influence the Max-Hold array content.
Setting
Command Description
SMHC
Max Hold
This command clears the content of the Analyzer Max-Hold array.
The reply is
MHC=OK
which acknowledges the command has been granted.
SMHF a
Max
Hold
Function
This setting command enables the Analyzer Max-Hold function according to the parameter
(a) which can be either
ON
or
OFF
.
The reply is
MHF=OK
which acknowledges the command has been granted.
When the command enables the Analyzer Max-Hold function it also disables the Analyzer
Max-Hold PAUSE making it thus immediately active.
#SMHF ON*.
Analyzer Max-Hold function is enabled and the sent spectrum reflects the
maximum value recorded, frequency by frequency, since it was last cleared.
#SMHF OFF*.
Analyzer Max-Hold function is disabled and the sent spectrum reflects the
current one.
SMHP a
Max
Hold
Pause
This setting command enables/disables the Analyzer Max-Hold Pause according to the
parameter (a) which can be either
ON
or
OFF
.
The reply is
MHP=OK
which acknowledges the command has been granted.
When the Pause is ON no updates are made on the max-hold array.
#SMHP ON*.
Analyzer Max-Hold function is suspended: all incoming values will not
influence the max-hold array contents.
# SMHP OFF*.
Analyzer Max-Hold function is enabled again: all max-hold array values
are updated to the maximum value.
Summary of Contents for PMM 9010 Series
Page 24: ...XXIV Safety considerations This page has been left blank intentionally ...
Page 42: ...2 10 Installation This page has been left blank intentionally ...
Page 66: ...6 6 Manual Mode Operating Instructions This page has been left blank intentionally ...
Page 72: ...7 6 Applications This page has been left blank intentionally ...
Page 92: ...9 14 General Information This page has been left blank intentionally ...
Page 112: ...10 20 Click Mode Operating Instructions This page has been left blank intentionally ...
Page 145: ...PMM 9030 9060 9180 Frequency Extension 11 33 Screwing the N N Adapter to BL 01 connector ...
Page 170: ...11 58 PMM 9030 9060 9180 Frequency Extension This page has been left blank intentionally ...
Page 178: ...12 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally ...
Page 194: ...13 16 APD Mode Operating Instructions This page has been intentionally left blank ...
Page 230: ...14 36 Remote control This page has been intentionally left blank ...
Page 284: ...Suggerimenti Commenti Note Suggestions Comments Note ...