14-24
Remote control
SPLM a
This setting command switches the Pulse Limiter; according the parameter (a) which
can be either
On
or
OFF
.
The reply is
PLM
=OK
which acknowledges the command has been granted or
PLM
=SERR
if the command has been ignored.
Example:
# PLM
ON*
Example:
# PLM
OFF*
SRBW b
This setting command sets RBW. The argument (b) should be a string representing
the RBW as follows:
o
0
RBW is Automatic
so that the 9010 automatically selects the
suitable RBW according to the tuned frequency
o
1
300kHz
o
2
100kHz
o
3
30kHz
o
4
10kHz
o
5
3kHz
o
Only for 9010
6
9kHz (Cispr 16-1)
7
200Hz (Cispr 16-1)
11
10kHz (MIL-STD-461) (only when available)
22
1kHz (MIL-STD-461) (only when available)
23
100Hz (MIL-STD-461) (only when available)
24
10Hz (MIL-STD-461) (only when available)
o
Only for Downconverter
9
1MHz (Cispr 16-1)
10
120kHz (Cispr 16-1)
12
9kHz (Cispr 16-1) ; No QPeak available
The reply is
RBW =OK
which acknowledges the command has been granted or
RBW =SERR
if the command has been ignored.
Example:
#S RBW 3*
sets the RBW to 30kHz and disables Auto RBW mode
Example:
#S RBW 0*
turn the RBW mode in Automatic
SSRT f
This setting command sets the Start frequency for Sweep mode. The string (f) can
be in exponential form and should be expressed in Hz. The reply is
SRT=OK
which
acknowledges the command has been granted or
SRT =SERR
if the command has
been ignored.
Example:
#SSRT 150e3*
Sets the Start frequency, for Sweep mode, to 150kHz
SSTP
This setting command stops abruptly a running sweep in Analyzer mode. The reply
is always
STP=OK,
which acknowledges the command has been granted
STAT b
This setting command sets MinAtt to the value indicated by (b) which range from
0dB to the maximum the 9010/9030/9060/9180 can set and must be in step of 5dB.
The reply is
TAT=OK
which acknowledges the command has been granted or
TAT
=SERR
if the command has been ignored.
Note: In Sweep mode the Attenuator is always forced to AUTO
Example:
#STAT 10*
STGF f
This setting command sets the tuning frequency for 9010 Built-In Generator when is
not tracking (STGT OFF
)
. The string (f) can be in exponential form and should be
expressed in Hz.
Frequency should range between 1 kHz and 50MHz inclusive.
The reply is
TGF=OK
which acknowledges the command has been granted or
TGF
=SERR
if the command has been ignored.
Example:
#STGF 150e3*
Sets the tuning frequency, for 9010 Built-In Generator, to
150kHz
Summary of Contents for PMM 9010 Series
Page 24: ...XXIV Safety considerations This page has been left blank intentionally ...
Page 42: ...2 10 Installation This page has been left blank intentionally ...
Page 66: ...6 6 Manual Mode Operating Instructions This page has been left blank intentionally ...
Page 72: ...7 6 Applications This page has been left blank intentionally ...
Page 92: ...9 14 General Information This page has been left blank intentionally ...
Page 112: ...10 20 Click Mode Operating Instructions This page has been left blank intentionally ...
Page 145: ...PMM 9030 9060 9180 Frequency Extension 11 33 Screwing the N N Adapter to BL 01 connector ...
Page 170: ...11 58 PMM 9030 9060 9180 Frequency Extension This page has been left blank intentionally ...
Page 178: ...12 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally ...
Page 194: ...13 16 APD Mode Operating Instructions This page has been intentionally left blank ...
Page 230: ...14 36 Remote control This page has been intentionally left blank ...
Page 284: ...Suggerimenti Commenti Note Suggestions Comments Note ...