Remote control
14-7
?DET
This query command
#?DET*
sends back a string reporting the five detector for
manual mode. Unit is fixed dBµV. Each value is separated by a semicolon. The
order is Peak; Qpeak; RMS;AVG;C-RMS:C-AVG
In case Qpeak is not available (RBW other than 200Hz, 9kHz [only for Band B]
,120kHz) Qpeak field is replaced by the string “----“.
In case C-RMS or C-AVG is not available (RBW other than 200Hz,
9kHz,120kHz,1MHz) respective field is replaced by the string “----“.
An additional string “OVER;” is appended at the end whenever an over range
condition occurred
Example of reply
: DET=23.22;17.09;16.23;11.36,16.01,15.50;
(which means Peak=
23.22
; Qpeak=
17.09
; RMS=
16.23
;AVG=
11.36
;C-
RMS=
16.01;
C-Avg
=15.50
Example of reply (No Cispr filter selected, over range shown)
:
DET=17.20;-----;11.98;9.57;-----;-----;OVER;
?DMD
This query command
#?DMD*
sends back a string expressing the status of
demodulator. The reply is either:
•
DMD=Off
(The Demodulator is OFF)
•
DMD =AM
(The Demodulator is On and demodulating AM )
?DMV
This query command
#?DMV*
sends back a string expressing the volume of
demodulator. The reply is made of the string
“DMV=”
followed by the magnitude of
volume ranging from 0 to 100
•
DMV=50
(The Demodulator Volume is 50)
?DPY
This query command
#?DPY*
sends back a the bitmap of display hardcopy
?FSA
This query command
#?FSA*
sends back a string expressing the number of
Sweep Records available. The reply: FSA= n: bands(1),Detector(1);
bands(2),Detector(2);………..; bands(n),Detector(n)
Where
•
n
is the number of records available.
•
Band (k) expresses the bands (or part of them) present in the sweep. k
varies from 1 to n.
o
0x01 --> Band A
o
0x02 --> Band B
o
0x04 --> Band C
o
0x08 --> Band D
o
0x10 --> Band E
•
Detector (k) expresses the Detector used t in the sweep. k varies from 1 to
n
o
0x0001 --> Peak
o
0x0002 --> Avg
o
0x0004 --> RMS
o
0x0008 --> QPeak
o
0x0010 --> C-Rms
o
0x0020 --> C-Avg
o
0x0200 --> Smart Avg
o
0x0400 --> Smart RMS
o
0x0800 --> Smart QPeak
o
0x1000 --> Smart C-Rms
o
0x2000 --> Smart C-Avg
In case no records are available the reply is:
FSA= N/A
Example of reply:
FSA= 4:1,2049;3,2049;2,2049;1,1;
Summary of Contents for PMM 9010 Series
Page 24: ...XXIV Safety considerations This page has been left blank intentionally ...
Page 42: ...2 10 Installation This page has been left blank intentionally ...
Page 66: ...6 6 Manual Mode Operating Instructions This page has been left blank intentionally ...
Page 72: ...7 6 Applications This page has been left blank intentionally ...
Page 92: ...9 14 General Information This page has been left blank intentionally ...
Page 112: ...10 20 Click Mode Operating Instructions This page has been left blank intentionally ...
Page 145: ...PMM 9030 9060 9180 Frequency Extension 11 33 Screwing the N N Adapter to BL 01 connector ...
Page 170: ...11 58 PMM 9030 9060 9180 Frequency Extension This page has been left blank intentionally ...
Page 178: ...12 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally ...
Page 194: ...13 16 APD Mode Operating Instructions This page has been intentionally left blank ...
Page 230: ...14 36 Remote control This page has been intentionally left blank ...
Page 284: ...Suggerimenti Commenti Note Suggestions Comments Note ...