14-16
Remote control
SLIW n, freq;
lev
This setting command is used for setting the custom limit frequency by frequency. It
consists of three fields as follows:
•
n
Is the index of frequency being written and it should range from 0 to 31
•
freq
Is the frequency, expressed in Hz (exponential notation allowed), being
written
•
lev
Is the level of the limit and is expressed in dBµV
The procedure to set a custom limit on PMM9010 is the following:
•
Send as many commands as the frequency points of limit are in upwards
order.
•
Terminate (and make it active) by sending the command
SLIE
Here is an example how to make a custom limit on PMM9010:
•
#SLIW 0
, 150e3; 66 * Send the first line 150kHz 66 dBµV
•
#SLIW 1
, 500e3; 56 * Send the second line 500kHz 56 dBµV
•
#SLIW 2
, 5e6; 56 * Send the third line 5MHz 56 dBµV
•
#SLIW 3
, 5e6; 60 * Send the fourth line 5MHz 60 dBµV
•
#SLIW 4
, 30e6; 60 * Send the fifth line 30MHz 60 dBµV
•
#SLIE Custom CISPR 14-1 * Send the name and make the limit active
Note that all the higher indexes in the limit are cleared. Therefore, a limit must be
made in upwards order otherwise all point, having a higher index previously written,
are cleared.
It is
responsibility of the user to make sure
the data sent are correct and
coherent.
The reply is
SLIW =OK
which acknowledges the command has been granted or
LIW
=SERR
if the command has been ignored.
SMAF f
This setting command sets the tuning frequency for manual mode. The string (f) can
be in exponential form and should be expressed in Hz. The reply is
MAF=OK
which
acknowledges the command has been granted or
MAF=SERR
if the command has
been ignored.
When the command is granted the 9010 goes to manual mode automatically
Example:
#SMAF150e3*
Sets the manual frequency to 150kHz
SMAN
This setting command puts the 9010 in manual mode. The reply is always
MAN=OK
which acknowledges the command has been granted.
The 9010 goes to manual mode.
Example:
#SMAN *
SMANP
This setting command puts the 9010 in manual mode and sets the default value to
the following settings:
•
RBW
Automatic
•
Attenuator
Automatic
•
MinAtt=10
dB
•
Preselector
On
•
Preamplifier
OFF
•
Hold Time =1000ms
•
Pulse Limiter = OFF
•
TG=OFF
The reply is always
MANP=OK
which acknowledges the command has been
granted.
The 9010 goes to manual mode.
Example:
#SMANP*
Summary of Contents for PMM 9010 Series
Page 24: ...XXIV Safety considerations This page has been left blank intentionally ...
Page 42: ...2 10 Installation This page has been left blank intentionally ...
Page 66: ...6 6 Manual Mode Operating Instructions This page has been left blank intentionally ...
Page 72: ...7 6 Applications This page has been left blank intentionally ...
Page 92: ...9 14 General Information This page has been left blank intentionally ...
Page 112: ...10 20 Click Mode Operating Instructions This page has been left blank intentionally ...
Page 145: ...PMM 9030 9060 9180 Frequency Extension 11 33 Screwing the N N Adapter to BL 01 connector ...
Page 170: ...11 58 PMM 9030 9060 9180 Frequency Extension This page has been left blank intentionally ...
Page 178: ...12 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally ...
Page 194: ...13 16 APD Mode Operating Instructions This page has been intentionally left blank ...
Page 230: ...14 36 Remote control This page has been intentionally left blank ...
Page 284: ...Suggerimenti Commenti Note Suggestions Comments Note ...