14-12
Remote control
SAFF start,stop
This setting command sets both Start and Stop frequency for Analyzer mode. The
strings (Start and Stop) can be in exponential form and should be expressed in Hz.
The reply is
AFF=OK
which acknowledges the command has been granted or
AFF
=SERR
if the command has been ignored.
Example:
#SAFF 150e3,30e6*
Sets the Start frequency to 150kHz and Stop
frequency to 30MHz for Analyzer mode
SAGO
This setting command puts the 9010 in Analyzer mode. The reply is always
SAGO=OK
, which acknowledges the command has been granted, followed by an
array, variable in size, representing the reply. Please refer to “
Analyzer Reply
”
section for details of the response.
The 9010 goes to Analyzer mode.
The 9010 goes to Analyzer mode.
Example:
#SAGO*
SAHT h
This setting command sets hold time for Analyzer mode. The string (h) should be
expressed in ms. The reply is
AHT=OK
which acknowledges the command has
been granted or
AHT =SERR
if the command has been ignored.
Example:
#S AHT 1500*
Sets the hold time, for Analyzer mode, to 1500 ms (or
1.5s)
SAOP f
This setting command sets the Stop frequency for Analyzer mode. The string (f) can
be in exponential form and should be expressed in Hz. The reply is
AOP=OK
which
acknowledges the command has been granted or
AOP =SERR
if the command has
been ignored.
Example:
#SAOP 10e6*
Sets the Stop frequency, for Analyzer mode, to 10MHz
SAPA a
This setting command switches Preamplifier, for Analyzer Mode, according the
parameter (a) which can be either
On
or
OFF.
The reply is APA=OK which acknowledges the command has been granted or
APA
=SERR
if the command has been ignored.
Example:
#S APA ON*
Example:
#S APA OFF*
SAPS a
This setting command switches Preselector, for Analyzer Mode, according the
parameter (a) which can be either
On
or
OFF
.
The reply is
APS=OK
which acknowledges the command has been granted or
APS
=SERR
if the command has been ignored.
Example:
# APS ON*
Example:
# APS OFF*
SART f
This setting command sets the Start frequency for Analyzer mode. The string (f) can
be in exponential form and should be expressed in Hz. The reply is
ART=OK
which
acknowledges the command has been granted or
ART =SERR
if the command has
been ignored.
Example:
#SART 150e3*
Sets the Start frequency, for Analyzer mode, to 150kHz
SCFA n
This setting command activates the conversion factor stored in the 9010 (if any)
where n is the ID of the conversion factor as shown in 9010 menu.
If
n
is not in the range 1 to 4 no conversion factor is active
The reply is:
CFA=OK (OFF)
in case there is not active conversion factor
CFA=OK
in case the conversion factor is active
Summary of Contents for PMM 9010 Series
Page 24: ...XXIV Safety considerations This page has been left blank intentionally ...
Page 42: ...2 10 Installation This page has been left blank intentionally ...
Page 66: ...6 6 Manual Mode Operating Instructions This page has been left blank intentionally ...
Page 72: ...7 6 Applications This page has been left blank intentionally ...
Page 92: ...9 14 General Information This page has been left blank intentionally ...
Page 112: ...10 20 Click Mode Operating Instructions This page has been left blank intentionally ...
Page 145: ...PMM 9030 9060 9180 Frequency Extension 11 33 Screwing the N N Adapter to BL 01 connector ...
Page 170: ...11 58 PMM 9030 9060 9180 Frequency Extension This page has been left blank intentionally ...
Page 178: ...12 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally ...
Page 194: ...13 16 APD Mode Operating Instructions This page has been intentionally left blank ...
Page 230: ...14 36 Remote control This page has been intentionally left blank ...
Page 284: ...Suggerimenti Commenti Note Suggestions Comments Note ...