11-52
PMM 9030/9060/9180 Frequency Extension
11.17.3.2 Misc
Under the
Miscellaneous
functions menu it is possible to activate or
exclude the
Preselector
filters and also to set the
Tracking generator
.
The
Preselector
is composed by a group of filters automatically selected
by the PMM 9030/9060/9180 while it is sweeping or anyway measuring.
The aim of the preselector is to reduce the amount of out-of-band energy
entering in the receiver, thus helping a lot in reducing intermodulation
problems and similar undesired behaviors.
It can be set either ON or OFF with the associated button, and normally it
should be always enabled.
On the top left corner of the screen the symbol “Off” or “On” will be
displayed.
Pressing the
RF OUT
button the receiver enters the
Tracking generator
menu.
The
Tracking generator
is an internal, high stability and accuracy, 50
Ohm RF generator ranging from 10 Hz to 50 MHz.
It is a CW signal source tuned at the frequency set under the
RF OUT
Freq
window.
As usual, to set a given frequency it is therefore enough to edit the value
with the soft keys and the left an right arrows, confirming the selections
entering the units value (kHz, MHz or GHz).
Pressing 0 as the first figure the decimal dot appears automatically.
The output level can be set between 60,0 and 90,0 dBµV with 0,1dB steps
using the
RF OUT
Level button.
If a higher or lower level is needed, the User shall adopt either an external
amplifier or an external attenuator.
With the PMM 9030/9060/9180 the
Tracking On
function is not available.
11.17.4 RBW
The
Resolution Bandwidth
command is used to select the bandwidth of
the measuring filter. Seven bandwidth filters are available:
•
120 kHz CISPR 16
•
3 kHz
•
10 kHz
•
30 kHz
•
100 kHz
•
300 kHz
•
1 MHz (Bimp)
The four larger filters are selectable from the first RBW screen, and
pressing the
More RBW
button it is possible to select all the other filters.
These filters are mathematically modeled using FIR (Finite Impulse
Response) technique and they are exactly as required by the standards.
More filters will be available as an option for specific applications, e.g. to
cover Military Standards requirements, etc.
With the
Auto Cispr
function, the filter will be automatically selected,
according to CISPR standard, depending on the tuned frequency.
When a non-CISPR filter is selected, the Quasi Peak detector is disabled.
Summary of Contents for PMM 9010 Series
Page 24: ...XXIV Safety considerations This page has been left blank intentionally ...
Page 42: ...2 10 Installation This page has been left blank intentionally ...
Page 66: ...6 6 Manual Mode Operating Instructions This page has been left blank intentionally ...
Page 72: ...7 6 Applications This page has been left blank intentionally ...
Page 92: ...9 14 General Information This page has been left blank intentionally ...
Page 112: ...10 20 Click Mode Operating Instructions This page has been left blank intentionally ...
Page 145: ...PMM 9030 9060 9180 Frequency Extension 11 33 Screwing the N N Adapter to BL 01 connector ...
Page 170: ...11 58 PMM 9030 9060 9180 Frequency Extension This page has been left blank intentionally ...
Page 178: ...12 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally ...
Page 194: ...13 16 APD Mode Operating Instructions This page has been intentionally left blank ...
Page 230: ...14 36 Remote control This page has been intentionally left blank ...
Page 284: ...Suggerimenti Commenti Note Suggestions Comments Note ...