MH-50B 

Semi-Auto

 

Chemistry Analyzer 

 

 

 

 

 

USER MANUAL 

 
 

 

Add: No.5 Hepanxiang

Jinhua

Zhejiang, China 321001 

Tel:  86-579-

 

82373070        Fax:  86-579-83533099 

Website: 

www.mossinstruments.com

   

 

Summary of Contents for MH-50B

Page 1: ...MH 50B Semi Auto Chemistry Analyzer USER MANUAL Add No 5 Hepanxiang Jinhua Zhejiang China 321001 Tel 86 579 82373070 Fax 86 579 83533099 Website www mossinstruments com...

Page 2: ...e 7 3 4 Test Example 10 3 5 RESULT REPORT 12 Chapter 4 Programme 16 Chapter 5 Q C and Reference Value 23 5 1 Preparation 23 5 2 Test Q C material 23 5 3 Print the curve and result of Q C 23 Chapter 6...

Page 3: ...which goes through the liquid layer so we can calculate the consistency of the given solution 1 2 Technical Criterions Use Condition Temperature 0 40 o C Relative humidity 10 90 Applied elec Power In...

Page 4: ...p 1 3 4 Menu Interface MEASURE PRINT REPORT MH 50B CHEMISTRY ANALYZER Ver X XX Inner thermal printer Display Screen Operation panel board ASP Key Digit Keys Cursor Keys FEED PAPER Waste Outlet Menu ke...

Page 5: ...instrument must be put in a platform which has enough room stable level without any vibration the situation should avoid the equipments such as fans air conditioners etc in order not to aeffect the ac...

Page 6: ...MH 50B Chemistry Analyzer Operation Manual 5 9 Follow the English introduction to operate...

Page 7: ...ANDARD liquor Absorb sample liquor and analyze it Press 8 to select the patient s No and print the result Press MENU back to main menu Press 3 enter the menu RESULT REPORT Press 1 to edit patient s in...

Page 8: ...lt press the digit key 2 in the main interface then input the given patient s number to print also you can press the digit number 8 in the Test items interface then input the given patient s number an...

Page 9: ...ance It will prompt you to absorb DI water after filter wheel rotated in order to adjust absorbance 3 Absorb clean liquor When you test the item for the first time you need to do Blank Test and Standa...

Page 10: ...est when Absorb standard liquor twinkles continually The test result will be shown in the above chart the system will change to the next patient of the same item to test automatically You can do the f...

Page 11: ...after finishing this step 3 4 Test Example 3 4 1 Test ALT GPT with Kinetic method 1 Installing the parameters in the menu EDIT TEST according to the reagent manual as follows TEST CODE 1 TEST NAME ALT...

Page 12: ...T 1 STANDARD 1 STD 1 70 0 2 Select the LIVER ITEM in TEST SAMPLE interface press the to page down press digit key 1 to select the TOTAL PROTEIN The screen will show Absorb clean liquor then put the CL...

Page 13: ...esting the standard liquor press the digit key 7 to modify the patient s No put the corresponding patient s sample and press ENTER key to test the sample liquor Attention The edit item end point do no...

Page 14: ...has been saved in memory you can modify at freedom and press Menu to return you can not input the patient s telephone No and address But there is space in the RESULT REPORT for you to add 3 5 2 Delet...

Page 15: ...CODE in the second row input the item No then press ENTER key it will show the item name in the third row and show the result in the fourth row the cursor will stop at the back of the result it will...

Page 16: ...suggests you input the start date press ENTER key directly to print all results in internal memory then it will show Print all results in internal memory instantly with pressing Y or ENTER Return to S...

Page 17: ...de is less than 200 it is a combination test if it is more than 200 or equal to 200 Input the test code press ENTER key the test item name will be shown at the back of the name you can use keys to cha...

Page 18: ...There are 8 wavelengths to select 340nm 380nm 405nm 492nm 510nm 546nm 578nm 620nm 8 Install REAGENT BLANK and SAMPLE BLANK according to the reagent manual press ENTER key or key to enter into the next...

Page 19: ...NTER key to save or press N key The system will be back to this section s beginning state after input all needed condition now you can input or modify the next test item Attention a If the test method...

Page 20: ...by CHANGCHUN HUILI GPT End Color Method Reagent Manual See in Appendix 1 Enter into EDIT TEST interface in main menu 2 Press 1 to enter into TEST ITEM DEFINITION interface input 1 in TEST CODE You ca...

Page 21: ...IS 300 SAMPLPOINTS IS 2 15 General reagent is single standard if there is no special explain input 1 in STANDARD If you want to enter into next item input STANDARD 1 input 100 here 16 Press Y to assu...

Page 22: ...ndard to100 It will show the test result after absorbing the sample liquor one by one 7 NORMAL VALUE 8 50u L 8 TECHNICAL INDEX AND NOTICE EVENTS BLANK ABSORBANCE A510nm 0 3 CV 4 7 LINEARITY 300u Atten...

Page 23: ...E TEST TIME REACTI ON DIRECTI ON REAGENT VOL Ul SAMPL E VOL Ul FAC TOR ISP I II 340 Kin 37 30 30 40 500 50 176 8 Beckma n700 340 Kin 37 30 30 40 500 50 241 2 Attention Icterus patient s best ratio of...

Page 24: ...lease remember the No because it will be used in the follow procedures then test the prepared Q C material according to the method of testing sample 2 2 Enter RESULT REPORT submenu after finishing tes...

Page 25: ...nt the correct Q C curve and result 2 If there is not target value for the Q C material you can test it for 10 times continually to assure all items target values then calculate the MEAN and SD as the...

Page 26: ...ADC It shows the internal AD transfer parameter of the instrument repairmen use it in repairing 4 Printer and communication 4 1 Set Printer a Print Content Select 1 PRINT ALL RESULTS then to print rep...

Page 27: ...ion result will be shown at the last row of the report list after installing the consent doctor s autograph SAVE DOCTOR S NAME Input the doctor s autograph when it suggests Please input doctor s autog...

Page 28: ...7 MAINTENANCE 6 Edit patients file The internal memory will be effected because of some reason then it will save wrong data it will effect result report if without modification this function is as fol...

Page 29: ...NUTE SECOND XX XX AM MORNING PM AFTERNOON If it is 24 HOURS SYSTEM XX is blank b Date Format YEAR MONTH DATE c Press MENU if you don t modify the time and date don t press ENTER or it will effect the...

Page 30: ...cially remain thing in the cliff of colorimetric cell The tube should be full of DI water at last 3 Maintenance for every week Clean the tube with DI water after testing sample twice then absorb the C...

Page 31: ...ustor No Printing a No paper b Print paper is blocked c Thermal print paper is in reverse a change the new paper b install the paper again c install the paper again Fail Absorbance a Do not absorb DI...

Page 32: ...al side downwards Take the roll axis out upwards put the thermal side under the roll axis as the chart the paper s direction should be vertically with the one of the roll axis it avoids the printer pa...

Page 33: ...en press 3 enter ADC interface 3 Press 8 clean 4 If use alcohol discharge the residual alcohol in tube with pressing 8 change DI water twice 5 Press MENU to return to the main menu after cleaning Appe...

Page 34: ...r Mea sure state Resu lt type DAT A decima l check sum Fram e end CO DE FA 01 01 01 0 7 F 0 7F 0 7 F 80H 8FH D0 D3 0 7F each E0 E4 D7 0 FBH Standard absorbance or result absorbance BI TE 1 1 1 2 1 1 3...

Page 35: ...orbance to test no unit data is 3 N bites for N standards c Use no unit absorbance and unit consistence to test d Absorbance value turn the decimal to left for 3 bites e Consistence turn the data to l...

Page 36: ...P 3 GOT 28 K 4 GT 29 Na 5 TP 30 CO2CP 6 ALB 31 Fe 7 TBIL 32 HGB 8 DBIL 33 RBC 9 TTT 34 GRP 10 BUN 35 A G 11 CRE 36 UDBIL 12 UA 37 IGG 13 CO2 38 C3 14 NH3 39 IGM 15 GLU 40 CHE 16 TCHO 41 CK 17 HDLC 42...