MH-50B Chemistry Analyzer Operation Manual
12
INTERVAL TIME: 10S ASP TIME: 300, SAMPLE POINT: 2,
STANDARD No.: 5, STD 1: 0.0 STD 2: 50.0
STD 3: 109 STD 4: 209, STD 5: 406.
2) Select the “LIVER TEST” in the “TEST SAMPLE” interface, press the digit key ‘1’ to select “GPT”.
The screen will show “Absorb clean liquor”, put the CLEAN and press ‘ENTER’ key to start the clean
operation. Put the DI water to test absorbance after finishing operation.
3
)
The screen will show “Absorb blank liquor” after testing, put the “Blank liquor” and press the
“ENTER” key to assure the operation.
4)Now the screen shows “Absorb standard liquor 1” (Press the ‘STANDARD’ key to turn to “Absorb
standard liquor 1” interface if it shows “Absorb sample liquor” ), put the standard liquors 1-5
successively, then press ‘ENTER’ key to test the standard liquor to calculate the parameters.
5) The screen shows “Absorb sample liquor” again after testing the standard liquor, press the digit key
‘7’ to modify the patient’s No, put the corresponding patient’s sample and press ‘ENTER’ key to test
the sample liquor.
Attention: The edit item(end-point) do not need to test the standard liquor after the first time, also you
can retest the standard liquor to assure the parameters. It is better to test the blank liquor before
testing the item everyday to get accurate results.
3.5 RESULT REPORT
Input the digit key ‘3’ in the main menu to enter the “RESULT REPORT” menu as follows; you can do
the EDIT.MODIFY.PRINT.DELETE etc. operations of the patient’s result in this menu.
3.5.1 Edit the patient’s information
You can input each patient’s name, gender, age and blood type according to this function.
First, the screen will display:
You can input the serial number directly or use the ‘
↓
’ key to search for the patient’s No , then press
“ENTER” key, the system will show as follows if the No. you input is not existed in the internal
memory:
SAMPLES SUM: 0
SAMPLE NO:
1-EDIT SAMPLE
4-PRINT/DELETE
2-DEL RESULT
5-TRANSMIT
3-ADD RESULT
6-RECEIVE