MH-50B Chemistry Analyzer Operation Manual
7
3.2 Common operation
3.2.1
Switch the elec. Power on,and incubate the instrument for 15minutes at least;
3.2.2
Test sample: Prepare the sample and reagent according to the reagent manual, enter the first
item in main menu after its reaction time add to the requirement minutes,:‘1-Test sample’ to test the
given sample;
3.2.3
Result report: after testing all items, press ‘menu’ key to enter the main menu, and press digit
key ‘3’ to enter the third item: “Result Report”, select “1-edit patient’s info”, input each patient’s info.
3.2.4
Print Report: After editing all patients’ info, press ‘MENU’ back to the “Result Report” board,
press digit key ‘4’ to print or delete patients’ results; If you only need to print one patient’s result, press
the digit key ‘2’ in the main interface, then input the given patient’s number to print; also you can press
the digit number ‘8’ in the Test items’ interface, then input the given patient’s number and print
directly;
3.2.5
Clean the pipe: clean the tube with CLEAN after you finish operation;
3.2.6
Power off the instrument: Finish all procedures, then cut off the elec. power.
3.3 Test Sample
Press 1 to enter into the menu “Test sample” interface, the system will show:
Press ‘
↑
’’
↓
’’
←
’’
→
’ to select test, there are more items can be selected:
Press ‘ENTER’ key to enter the interface of “absorb clean liquor”.
1. Absorb the clean liquor
Enter the interface of “Absorb clean liquor”, the screen will display as follows:
Zn
P
K
Na
CO2CP
Fe
HGB RBC GRP A/G
UDBIL IGG
IGM C3
CHE
CK
BU
U-PRO U-Hb U-GLU
U-URO U-BIL U-KET U-NIT U-PH
PRESS ARROW TO SELECT TEST
ALT
ALP
GOT r-GT
TP
ALB Tbil
DBIL TTT
BUN
CRE UA
CO2 NH3 GLU
TCHO HDLC LDLC TG
LDH
AMS CL
Ca
Mg
Cu
PRESS ARROW TO SELECT TEST