MH-50B Chemistry Analyzer Operation Manual
2
Chapter 1 Introduction
1.1 Test principle
The Bio-chemical analyzer is mainly used for the quantitative analysis of the clinical bio-chemical
items, it applies to the Lamber- Beer law:
A=Kbc=lg(I
0
/I)
Among them:
A: Absorbance
K: Absorbent Index
B: Liquid layer’s Thickness
C: Solution’s Consistency
I
0
: Incident light’s Strength
I: Transmit light’s Strength
According to the above formulas, under the condition of the certain liquid layer’s thickness and
incident light’s strength, as long as test the strength of transmit light which goes through the liquid
layer , so we can calculate the consistency of the given solution.
1.2 Technical Criterions
Use Condition
:
Temperature: 0 – 40
o
C
;
Relative humidity
: 10 – 90 %
;
Applied elec. Power: Interflow 220V±10%
;
Conserved Transportation Condition:
Temperature: 0 – 40
o
C
;
Relative humidity
: 10 – 90 %
;
Amount of saved items’ parameters:
≤
70
Solidify items’ parameters in out-firm : 50 units, described largely in “Chapter 4”
The result numbers instrument can save: 200 patients or quality control
The result numbers each patient can save:
≤
30 units items
Absorbing samples volume:<600ul
Inner Printer: 36 bits thermal printer(wide:80mm)
(Printer paper: Thermic , 80mm
Screen: 240*128 dots graphic LCD
Communications connector: RS-232)
1.3 Structure Introduction
1.3.1 Main components view