MH-50B Chemistry Analyzer Operation Manual
17
(4). Select analyze method with ‘
→
’ ‘
←
’ keys at the back of ‘METHOD’. There are 6 methods to select:
E-ENDPOINT.K-KINEMIC.B-DOUBLE-WAVELENGTH.C-SUBTRACTION.D-DIVISION.H-YIN and
YANG OPPOSITES.
This instrument can do all above tests, so please care that if the reagent is fit for the instrument when
you buy the reagent. Press ‘
↓
’ or ‘ENTER’ key to enter into the next screen:
(5). This instrument has Temperature Control system, you can select “37°C” or “30°C” or
“ROOM-TEMP” state with ‘
←
’ ‘
→
’;
(6). Input the needed decimals at the back of ‘DECIMAL’, input 2 digits at most, input 0 if there is
no decimal;
(7). Select the needed wavelength of the test item at the back of ‘WAVELENGTH 1’ by ‘
→
’
‘
←
’ ; When the ‘ANALYZE METHOD’ select ‘DOUBLE-WAVELENGTH’, it will show ‘WAVELENGTH
2:’, There are 8 wavelengths to select: 340nm.380nm.405nm.492nm.510nm.546nm.578nm.620nm ;
(8). Install ‘REAGENT BLANK’ and ‘SAMPLE BLANK’ according to the reagent manual; press
‘ENTER’ key or ‘
↓
’ key to enter into the next item:
(9).You can install the item’s reference value, this instrument allows to input 4 different reference
values according to the clinic application, if you have installed the patient gender, the system will
transfer the corresponding reference value, or it will transfer the common reference normal value.
Then press ‘ENTER’ or ‘
↓
’ key to enter into next item.
(10). Find the reagent’s linear time in the reagent manual, select “<” or “>” with cursor key ‘
→
’or ‘
←
’,
press ‘ENTER’ key to enter into the next item, input number “200”.
1 ALT
LINEAR RANGE:< 600
DELAY TIME: 60S
INTERVAL TIME: 60S
DATA NO: 2
ASP TIME: 250
PRESS ‘
←
’ ‘
→
’ TO SELECT
1 ALT REFERENCE RANGE
GENERAL MALE FEMALE INFANT
HIGH: 40 40 40 40
LOW: 0 0 0 0
1 ALT
TEMP:37
℃
DECIMAL: 0
WAVELENGTH:340nm
REAGENT BLANK: Y SAMPLE BLANK: N
PRESS ‘
←
’ ‘
→
’ TO SELECT