MH-50B Chemistry Analyzer Operation Manual
19
Input item name directly and press ‘ENTER’ key to save. Press ‘MENU’ to exit.
Add names of combination and suppliers is close to add item name. You can select the added name
in the corresponding section of “TEST SETUP”.
3). Delete test name
You can delete the needless item name or item, If there is item in patient result, please delete it firstly,
or it will show wrong suggestion. It will show:
Press ‘ENTER’ to delete the current test name.
4). Delete test
You can delete the needless item, it will show as below after entering this interface:
Press cursor key to select the test, press “ENTER” to delete it.
5). Example of edit test
(1). Endpoint
For example: edit the test parameter of GPT according to GPT manual produced by
CHANGCHUN HUILI “GPT End Color Method Reagent Manual” (See in Appendix).
1. Enter into “EDIT TEST” interface in main menu.
2. Press ‘1’ to enter into “TEST ITEM DEFINITION” interface, input 1 in TEST CODE(You can look
up the solidified item amount and name of the instrument in item form , the item No. is 1).Press
‘ENTER’ key.
3. Select “GPT” with cursor keys ‘
←
’ ‘
→
’ in “NAME”, press ‘ENTER’ key to enter into next item.
4. Select “HUILI” WITH CURSOR KEYS ‘
←
’ ‘
→
’ IN “REAGENT SUPPLIERS”.
5. Input 0 in “COEFFICIENT”, do BLANK TEST and STANDARD TEST after editing, the system will
setting the value automatically.
6. Explain “NORMAL VALUE: 8—50u/L” in manual, you will know the reagent test unit is u/L. Search
the supplier with cursor keys ‘
←
’ ‘
→
’, press ‘ENTER’ key or ‘
↓
’ to turn to next item.
ALT
ALP
GOT r-GT
TP
ALB Tbil
DBIL TTT
BUN
CRE UA
CO2 NH3 GLU
TCHO HDLC LDLC TG
LDH
AMS CL
Ca
Mg
Cu
PRESS ARROW TO SELECT TEST
PRESS ‘ENTER’ TO DELETE
DELETE TEST NAME
ADDED NAME SUM: 2
THE NAME TO DELETE:EK
PRESS ‘
←
’ ‘
→
’ TO SELECT