MH-50B Chemistry Analyzer Operation Manual
34
Test code: (80-0xBF, 64 codes at most)
NUM
TEST CODE
CODE
NUM
TEST CODE
CODE
1 1 81H
8 8
88H
2 2 82H
9 9
89H
3 3 83H
10 10
8AH
4 4 84H
11 11
8BH
5 5 85H
…… …… ……
6 6 86H
…… …… ……
7 7 87H
…… …… ……
Unit code:
NUM NAME
CODE
NUM NAME
CODE
0 C0H
8
g/dL
C8H
1
u/L
C1H 9
g/mL
C9H
2
u/
m
L
C2H 10 L
eu/uL
CAH
3 u
mol/L C3H 11 X1012
/
L
CBH
4 m
mol/L C4H 12
CCH
5
g/L
C5H 13
CDH
6
mg/L
C6H 14
7
mg/dL
C7H 15
Measure State
Result Type
NUM NAME CODE
NUM NAME CODE
1 BLANK D0H 1 originality D3H
2 STANDARD D1H 2 Absorbance D4H
3 SAMPLE D2H 3
Consistence
D5H
Attention:
a)Use originality to test BLANK, no unit;
b)Use absorbance to test, no unit, data is 3*N bites for N standards;
c) Use no unit absorbance and unit consistence to test;
d) Absorbance value: turn the decimal to left for 3 bites;
e) Consistence: turn the data to left for (D
decimal
—0xE0)bites;.
f) The CRC check begins with the start bite of FA and end with front one bite of the check sum;
g) Item’s result data bite and patient No. are ((D
o
&7F)<<14)+((D
1
&7F)<<7)+(D
2
&7F);
h) The frame length is which begins with the item type bit and end with the front one bit of the check
sum.
i) The first frame’s No. is 1, and it will add 1 when send one frame till 0x7f, then begins with 1 again.
j) The instrument sends data to the serial when the screen shows the result report. For example, it will
send data automatically when get the result in “TEST SAMPLE” and “RESULT REVIEW”.