MH-50B Chemistry Analyzer Operation Manual
10
Press the digit key ‘8’, the screen will show as follows:
Input “Patient No.” and press “ENTER” key, the system will print the patient’s result report instantly,
press “MENU” key back to “Test Sample” interface;
Press the digit key ‘9’ to setting the item’s coefficients;
Press the ‘MENU’ key back to main menu.
If the item applies Kinetics method, the screen will show the reaction curves when you test the
standard and samples, such as follows:
Press “ENTER” key, the screen will show the recent curve to observe it better, press “MENU” key to
return. Or the screen will be back to “Absorb sample liquor” interface to check the result and test next
time after finishing this step
3.4 Test Example
3.4.1 Test ALT/GPT with Kinetic method
1) Installing the parameters in the menu “EDIT TEST” according to the reagent manual as follows:
TEST CODE: 1 , TEST NAME
:
ALT/GPT , REAGENT SUPPLIER: HUILI
FACTOR: 1746, UNIT: u / L , METHOD:DYNAMIC
TEMP: 37
o
C , DECIMALS: 0, WAVELENGTH: 340nm ,
REAGENT BLANK: N, SAMPLE BLANK: N HIGHT: 36.0,
LOW: 8.0, LINEARITY RANGE: < 500, DELAY TIME: 30S,
INTERVAL TIME: 40S ASP TIME: 300, SAMPLE POINTS: 2,
STANDARD: 0,
2) Input the digit ‘1’ in “Test Sample” interface. select ‘1-LIVER TEST’ then select the ALT,
the screen will show “Absorb clean liquor”, put the CLEAN
,
then press “ENTER” key to start the clean
operation.
3) The screen will show “Absorb sample liquor”, please assure the patient No., press the digit ‘7’ to
TEST CODE: 2 ALP
SAMPLE NO: 6
PRINT:OFF
SAMPLE NO:
WAVELENGTH: 340nm
ZERO: 4 1 STD: 100
ABSORBANCE: 0.154
RESULT: 37 mmol/L
SAMPLE: 6
NEW SAMPLE NO: