MH-50B Chemistry Analyzer Operation Manual
16
Chapter 4 Programme
This chapter mainly introduces how to set the reagent’s test parameters when you use it for the first
time. It explains that setting the parameters again according to the reagent manual after replacing the
reagent every time to get the right result.
Select ‘EDIT TEST’ in main menu to enter this function interface which includes 4 functions:
1). Test item definition
Its main function consists in inputting the operation parameters of every test item into the instrument
in order it can carry out the test function according to the correct instruction.
Dependent item definition
Press key ‘1’ to enter ‘1-TEST SETUP’, select the item, the system shows:
(1). Input a code of 1—3 bits in the cursor, it can not be 0, press ‘ENTER’ key if it is 1 bit or 2 bits, it is
a dependent test item if the code is less than 200, it is a combination test if it is more than 200 or
equal to 200. Input the test code, press ‘ENTER’ key, the test item name will be shown at the back of
the name, you can use ‘
→
’ ‘
←
’ keys to change the tests name; use ‘
→
’ ‘
←
’ keys to select the reagent
supplier, there are 9 consent suppliers(Test name and suppliers name can be modified .added or
deleted in the menu of “ADD NAME” and “DELETE NAME” , they can not be modified here)
Select the supplier’s name, press ‘ENTER’ key to enter the next interface:
(2). Now the cursor is at the back of the ‘COEFFICIENT’, you can modify or input the test
coefficient(factors) from ‘0—59999’,if you input the number from 1bit to 5 bits, please press ‘ENTER’
key or press ‘
↓
’ key directly to turn to the next item.
(3). Select unit with ‘
←
’ ‘
→
’ keys, press ‘
↓
’ or ‘ENTER’ key to enter into next item, there are 11 units to
select: None.u/L.u/mL.umol/L.mmol/L.g/L.mg/L.mg/Dl.g/ml.g/Dl.X10(12)/L.
1 ALT
FACTOR: 0
UNIT: u/L
METHOD:KINETICS
PRESS ‘
←
’ ‘
→
’ TO SELECT
TEST CODE: 1
TEST NAME:ALT
PRESS ‘
←
’ ‘
→
’ TO SELECT
1-TEST SETUP
2-ADD TEST NAME
3-DELETE TEST NAME
4-DELETE TEST