MH-50B Chemistry Analyzer Operation Manual
9
4. Absorb standard liquor 1
Put standard liquor, press “ENTER” key, the instrument starts to absorb clean liquor,it will test the
absorbance and original value of Blank liquor, the test result will be shown in the chart.
Press “CLEAN” key to clean again.
Press “BLANK” key to test the blank liquor again.
Press “STANDARD” key to retest standard liquor again.
Press “MENU” key back to main menu.
5. Absorb sample liquor
If the instrument ever do the Blank Test and Standard Test for the item, it will enter into “Sample Test”
interface after cleaning and testing absorbance just as the above chart. If it is the first time to test the
item, you should do the Blank Test and Standard Test before testing the samples.
Press the “ENTER” key to absorb the sample to do the test when “Absorb standard liquor” twinkles
continually. The test result will be shown in the above chart, the system will change to the next patient
of the same item to test automatically;
You can do the follow operations in this interface (When “Absorb sample liquor” twinkles continually)
Press ‘Standard’
,
‘Blank’
,
‘Clean' keys to do the same operation
Press the digit key‘6’ to enter the item and select interface to test other items:
Press the digit key‘7’to alter the patient No., just show as follows:
TEST CODE: 2 ALP
SAMPLE NO: 6
PRINT:OFF
ASP SAMPLE
WAVELENGTH: 340nm
ZERO: 4 1 STD: 100
ABSORBANCE: 0.154
RESULT: 37 mmol/L
TEST CODE: 2 ALP
SAMPLE NO:
PRINT:OFF
ASP STANDARD
WAVELENGTH: 340nm
ZERO: 4 1 STD: 100
STANDARD ABS: 0.723
FACTOR: 138.2
TEST CODE: 2 ALP
SAMPLE NO:
PRINT:OFF
ASP STANDARD
WAVELENGTH: 340nm
ZERO: 4 1 STD: 100
BLANK ABS: 0.000
ORIGINAL VALUE: 5445
Shows after finishing
Blank Test
Shows the first
patient no
automatically
Shows after
finishing the
Sample Test
Shows after finishing
Standard Test