MH-50B Chemistry Analyzer Operation Manual
28
handwork modification, or it will show wrong result.
Method: Press ‘1’, the cursor will appear in A, modify the value in A equal to the one in B then press
‘ENTER’ key to correct the mistake.
(2) If you delete the patient’s information in patients’ file, search the patient with ‘
↓
’ ‘
↑
’, the patient No.
will be shown in D, press ‘2’ to delete the patient.
Attention:
This operation can not be recovered, so be careful. If you delete the appointed result of
the patient, operate in “RESULT REPORT”.
(3) Press ‘MENU’ back to last menu.
7. Set the time
Enter “CLOCK SETUP” menu, it will show:
a>. Time Format: HOUR-MINUTE-SECOND XX: XX: AM: MORNING
PM: AFTERNOON
If it is 24 HOURS SYSTEM, XX is blank.
b>. Date Format: YEAR-MONTH-DATE
c>. Press ‘MENU’ if you don’t modify the time and date, don’t press ‘ENTER’, or it will effect the time’s
correctness.
8. System Reposition
When the system is disturbed by other reasons, this function can help you to reposit the system be
back to out-firm installation. Input ‘99999’ in ‘UTILITY’ to startup the function, it will ask you:
Press ‘1’ to reposit the system, then press ‘Y’ twice;
Press ‘2’ to clean the patients file, then press ‘Y’ twice;
Press ‘MENU’ to return.
All patients’ information and results will be lost after reposition in internal memory, so be careful.
PRESS ‘1’ TO RESET SYSTEM
PRESS ‘2’ TO RESET REPORTS
PRESS ‘MENU’ RETURN
HR FORMAT 1-12 Hr
√
2-24 Hr
DATE: 2008-01-01
TIME: 12:01:10
‘
←
’’
→
’CHANGE HR FORMAT
NUM KEY CHANGE TIME