MH-50B Chemistry Analyzer Operation Manual
24
according to the last Q.C. result.
Attention:
1.The operation for all patients in every test are the same except the test items. Select the
patient No. and the same test items for each test, you can not test other items for the appointed
patient No., or it will not print the correct Q.C. curve and result.
2. If there is not target value for the Q.C. material, you can test it for 10 times continually to assure all
items’ target values, then calculate the MEAN and SD as the target value and deviation. The target
value should be ascertained when you change new reagent.
3. Because the normal value range for different areas and different people are different, also the
instruments are different, other range only can be reference. Please creat the normal range fit to this
area’s people!