MH-50B Chemistry Analyzer Operation Manual
18
(11). Find the “DELAY TIME” and “TEST TIME” in reagent manual section of “TEST METHOD”, input
the times in the 2 positions. Do not modify the sampling time, it is only efficient for this test without any
influence for other tests. Standard is 0 for kinemics, standard is 1 for general reagent, you must install
the standard for multi-standard reagent; press ‘ENTER’ or ‘
↓
’ key to turn to next item after installing
the standard.
(12). You can install the item’s standard or standard curve in this interface.
Press ‘ENTER’ or ‘
↓
’ key to turn to next item after installing the standard. The system will suggest
‘Save modification?’ Press ‘Y’ or ‘ENTER’ key to save, or press ‘N’ key; The system will be back to
this section’s beginning state after input all needed condition, now you can input or modify the next
test item.
Attention: a. If the test method you selected is : SUBTRACTION, press ‘
↓
’ or ‘ENTER’ key, it will show:
Input all test codes under A.B.C.D.E.F, the corresponding codes for A.D must be existed, the decimal
of all tests must be equal, or it can not get correct result.
b. If the test method you selected is: DIVISION, press ‘ENTER’ or ‘
↓
’ it will show:
Input all test codes under A.B, the corresponding codes for A.B must be existed, the decimal of all
tests must be equal, or it can not get correct result.
2). Add new test name
You can add names of item combination or supplier for this operation, it will show:
ADD TEST NAME
ADDED NAME SUM: 0
PLEASE INPUT TEST NAME:
Y= A / B + C – D – E – F
Y= 2/ 46
Y= A + B + C – D – E – F
Y= 2 + 57 + 46 – 0 – 0 – 0