D - 4
SpO2 Low Signal
Quality
Low
C
1. Check the sensor and sensor
position.
2. Make sure the patient is not
shivering or moving.
3. The patient’s pulse may be too low
to be measured.
SpO2
Interference
Low
C
The SpO
2
signal has been interfered.
Check for any possible sources of
signal noise and check the patient
for excessive motion.
SpO2 Sensor
Error
Low
C
Replace the sensor and measure
again.
SpO2 Searching
Pulse
Prompt
/
SpO
2
is searching for pulse.
SpO2 Low
Perfusion
Prompt
/
The SpO
2
sensor is not properly
placed or the patient’s perfusion
index is too low.
1. Check the sensor and sensor
position.
2. Reposition the sensor if necessary.
Replace SpO2
Cable
Low
C
The SpO
2
cable expired, cannot be
recognized, or failed. Replace the
SpO
2
cable.
Prompt
The SpO
2
cable is about to expire.
Replace the SpO
2
cable.
Incompatible
SpO2 Cable
Low
C
The SpO
2
cable connected is not
compatible with the equipment.
Replace the SpO
2
cable.
Replace SpO2
Sensor
Low
C
The SpO
2
sensor expired. Replace
the SpO
2
sensor.
Prompt
The SpO
2
sensor is about to expire.
Replace the SpO
2
sensor.
Check SpO2
Sensor
Connection
Low
C
No SpO
2
sensor is connected.
Connect an SpO
2
sensor.
Alarm message
Default
priority
Indication on
alarm reset
Cause and solution
Summary of Contents for VS 8A
Page 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 8: ...2 6 This page intentionally left blank ...
Page 26: ...1 6 This page intentionally left blank ...
Page 58: ...3 24 This page intentionally left blank ...
Page 66: ...5 4 This page intentionally left blank ...
Page 70: ...6 4 This page intentionally left blank ...
Page 98: ...8 16 This page intentionally left blank ...
Page 142: ...13 4 This page intentionally left blank ...
Page 162: ...15 16 This page intentionally left blank ...
Page 166: ...16 4 This page intentionally left blank ...
Page 202: ...19 8 This page intentionally left blank ...
Page 224: ...22 10 This page intentionally left blank ...
Page 250: ...B 8 This page intentionally left blank ...
Page 260: ...C 10 This page intentionally left blank ...
Page 271: ...D 11 This page intentionally left blank ...
Page 272: ...D 12 ...
Page 275: ...F 1 F Declaration of Conformity ...
Page 276: ...F 2 This page intentionally left blank ...
Page 277: ......
Page 278: ...P N 046 020097 00 4 0 ...