D - 5
D.2.3
Temp Technical Alarm Messages
SpO2 No Cable
Connected
Low
C
No SpO
2
cable is connected. Connect
an SpO
2
cable.
Alarm message
Default
priority
Indication on
alarm reset
Cause and solution
Alarm message
Default
priority
Indication on
alarm reset
Cause and solution
Cannot Take
Neonatal Rectal
Temp (for
SmarTemp)
High
B
Rectal measurement is not
applicable for neonatal patient.
Select another site.
Temp No Probe
(for SmarTemp)
Med
B
The Temp probe is disconnected or
damaged. Reconnect the probe. If
the problem persists, change for
another probe.
Temp Probe
Misplaced (for
SmarTemp™)
Med
A
The measured value is invalid.
Replace the probe and check the
value.
Temp Measuring
Timeout (for
SmarTemp™)
Med
B
The measurement has exceeded
regular measurement time. Try
another measurement.
Ambient Temp
Overrange (for
SmarTemp™)
Low
A
The ambient temperature is too low.
Move the patient to a warmer place
and take temperature measurement
again if necessary.
Temp Overrange
(for
SmarTemp™)
Low
C
The temperature measurement
exceeds the measurement range.
Check the patient’s condition.
Thermal Resistor
Err (for
SmarTemp™)
Med
C
The Temp module may fail. Take
temperature measurement again. If
the alarm persists, contact your
service personnel.
TempIF
Thermometer
Error (for
Genius™ 3)
High
C
The thermometer may fail. Take
temperature measurement again. If
the alarm persists, replace the
thermometer.
Summary of Contents for VS 8A
Page 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 8: ...2 6 This page intentionally left blank ...
Page 26: ...1 6 This page intentionally left blank ...
Page 58: ...3 24 This page intentionally left blank ...
Page 66: ...5 4 This page intentionally left blank ...
Page 70: ...6 4 This page intentionally left blank ...
Page 98: ...8 16 This page intentionally left blank ...
Page 142: ...13 4 This page intentionally left blank ...
Page 162: ...15 16 This page intentionally left blank ...
Page 166: ...16 4 This page intentionally left blank ...
Page 202: ...19 8 This page intentionally left blank ...
Page 224: ...22 10 This page intentionally left blank ...
Page 250: ...B 8 This page intentionally left blank ...
Page 260: ...C 10 This page intentionally left blank ...
Page 271: ...D 11 This page intentionally left blank ...
Page 272: ...D 12 ...
Page 275: ...F 1 F Declaration of Conformity ...
Page 276: ...F 2 This page intentionally left blank ...
Page 277: ......
Page 278: ...P N 046 020097 00 4 0 ...