12 - 4
4.
Connect the CO
2
gas outlet to the scavenging system with an exhaust tube.
CO
2
can be measured after the start-up is complete.
NOTE
•
When sample gas of 37 °C, sample flowrate of 50 ml/min, room temperature
of 23 °C, 100% RH, the sampling line with a general type should be replaced
once at most every 8 hours, and the sampling line with a humidified type
should be replaced once at most every 72 hours.
•
If not necessary, do not disconnect the CO
2
adapter from the equipment after
the first installation.This reduces the risk of the CO
2
adapter becoming lost or
damaged.
12.6
Automatic CO
2
Module Zeroing
The CO
2
module performs zero calibration automatically when needed.
NOTE
•
The CO
2
module temporally stops measuring during zeroing.
12.7
Changing CO
2
Settings
12.7.1
Changing CO
2
Alarm Settings
In
Continuous Monitoring
mode, you can change the CO
2
alarm settings. Follow this
procedure:
1.
Select the CO
2
numeric area or waveform area to enter the
CO2
menu.
2.
Select
Alarm
tab.
3.
Set the following alarm properties:
◆
Switch on or switch off the alarms.
◆
Adjust the alarm limits and alarm priority.
Summary of Contents for VS 8A
Page 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 8: ...2 6 This page intentionally left blank ...
Page 26: ...1 6 This page intentionally left blank ...
Page 58: ...3 24 This page intentionally left blank ...
Page 66: ...5 4 This page intentionally left blank ...
Page 70: ...6 4 This page intentionally left blank ...
Page 98: ...8 16 This page intentionally left blank ...
Page 142: ...13 4 This page intentionally left blank ...
Page 162: ...15 16 This page intentionally left blank ...
Page 166: ...16 4 This page intentionally left blank ...
Page 202: ...19 8 This page intentionally left blank ...
Page 224: ...22 10 This page intentionally left blank ...
Page 250: ...B 8 This page intentionally left blank ...
Page 260: ...C 10 This page intentionally left blank ...
Page 271: ...D 11 This page intentionally left blank ...
Page 272: ...D 12 ...
Page 275: ...F 1 F Declaration of Conformity ...
Page 276: ...F 2 This page intentionally left blank ...
Page 277: ......
Page 278: ...P N 046 020097 00 4 0 ...