11 - 10
11.8.5
Setting the NIBP Display Format
To set the NIBP display format, follow this procedure:
1.
In
Continuous Monitoring
mode, select the NIBP numeric area to enter the
NIBP
menu. In
Spot Check
mode, select the
Main Menu
quick key
→
from the
Parameters
column select
Setup
→
select
NIBP
.
2.
Select the
Setup
tab.
3.
Set
Display Format
.
11.8.6
Setting the NIBP Alarm Limits Display Switch
In
Continuous Monitoring
mode, you can set whether to display the alarm limits of
diastolic NIBP and mean NIBP. Follow this procedure:
1.
Select the NIBP numeric area to enter the
NIBP
menu.
2.
Select the
Setup
tab.
3.
Switch on or off
Display Alarm Limits.
11.8.7
Selecting NIBP Measurement Algorithm
You can change the NIBP measurement algorithm adopted on your monitor. Follow this
procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required password
→
select
→
Module
.
2.
Under the
NIBP
tab, set
NIBP Measurement Algorithm
to:
◆
Inflation
: the monitor measures blood pressure during cuff inflation. If
succeeded, this measurement is completed. If the measurement fails at the
inflation stage, the monitor will automatically continue the measurement at
deflation stage.
◆
Deflation
: the monitor measures blood pressure during cuff deflation.
Affected by measurement factors or patient’s physiological conditions, inflation
algorithm might fail, be incorrect, or unavailable. Then the monitor automatically
continues to perform the NIBP measurement at deflation stage. The measurement
factors or patient’s physiological conditions include:
■
Excessive patient movement
■
Using cuffs with no
symbol
■
Using cuffs of inappropriate size
■
Cuffs being too tight or loose
■
Using well-worn cuffs
■
Patient blood pressure being too low
■
Patient pulse rate being too low
Summary of Contents for VS 8A
Page 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 8: ...2 6 This page intentionally left blank ...
Page 26: ...1 6 This page intentionally left blank ...
Page 58: ...3 24 This page intentionally left blank ...
Page 66: ...5 4 This page intentionally left blank ...
Page 70: ...6 4 This page intentionally left blank ...
Page 98: ...8 16 This page intentionally left blank ...
Page 142: ...13 4 This page intentionally left blank ...
Page 162: ...15 16 This page intentionally left blank ...
Page 166: ...16 4 This page intentionally left blank ...
Page 202: ...19 8 This page intentionally left blank ...
Page 224: ...22 10 This page intentionally left blank ...
Page 250: ...B 8 This page intentionally left blank ...
Page 260: ...C 10 This page intentionally left blank ...
Page 271: ...D 11 This page intentionally left blank ...
Page 272: ...D 12 ...
Page 275: ...F 1 F Declaration of Conformity ...
Page 276: ...F 2 This page intentionally left blank ...
Page 277: ......
Page 278: ...P N 046 020097 00 4 0 ...