7 - 5
You can also change the parameter alarm properties based on the condition of the
current patient. To do so, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
Alarm
column select
Limits.
Enter
the password if required.
2.
Select a parameter tab and set alarm properties as desired.
You can also change the alarm properties of individual parameter from corresponding
parameter menu.
The alarm settings of workflow override the alarm settings under
Limits
menu and
parameter menus. Every time you discharge a patient or switch to another workflow, the
workflow settings are loaded.
WARNING
•
When monitoring patients that are not continuously attended by a clinical
operator, properly configure the alarm system and adjust alarm settings as
per the patient's condition.
7.6.2
Setting Alarm Tone Properties
7.6.2.1
Changing the Alarm Volume
To change the alarm volume, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
Alarm
column select
Setup.
2.
Set
Alarm Volume
. The optional alarm volume is between X to 10, in which X is
the minimum volume, depending on the setting of minimum alarm volume, and
10 is the maximum volume.
3.
Select
High Alarm Volume
to set the volume of the high priority alarm.
4.
Select
Reminder Volume
to set the volume of the reminder tone.
NOTE
•
When the alarm volume is set to 0, the alarm sound is turned off and the
audio off symbol appears on the screen.
•
You cannot set the volume of high priority alarms if Alarm Volume is set to 0.
7.6.2.2
Setting the Interval between Alarm Sounds
If you choose the ISO pattern, you can change the interval between alarm tones. To
change the interval between alarm tones, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required password
→
select .
Summary of Contents for VS 8A
Page 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 8: ...2 6 This page intentionally left blank ...
Page 26: ...1 6 This page intentionally left blank ...
Page 58: ...3 24 This page intentionally left blank ...
Page 66: ...5 4 This page intentionally left blank ...
Page 70: ...6 4 This page intentionally left blank ...
Page 98: ...8 16 This page intentionally left blank ...
Page 142: ...13 4 This page intentionally left blank ...
Page 162: ...15 16 This page intentionally left blank ...
Page 166: ...16 4 This page intentionally left blank ...
Page 202: ...19 8 This page intentionally left blank ...
Page 224: ...22 10 This page intentionally left blank ...
Page 250: ...B 8 This page intentionally left blank ...
Page 260: ...C 10 This page intentionally left blank ...
Page 271: ...D 11 This page intentionally left blank ...
Page 272: ...D 12 ...
Page 275: ...F 1 F Declaration of Conformity ...
Page 276: ...F 2 This page intentionally left blank ...
Page 277: ......
Page 278: ...P N 046 020097 00 4 0 ...