18 - 10
18.6.5
The Other Tab
Alarm Priority
High Only
Selects the priority of alarms sent to the nurse call
system
Alarm Type
Physiological
Only
Selects the type of alarms sent to the nurse call
system.
Menu Item
Default Setting
Function
Section
Menu Item
Default
Setting
Function
Alarm Priority
SpO2 Sensor
Off
Low
Selects the alarm level for SpO
2
sensor off
alarm.
CMS/eGW
Disconnected
Low
Selects the priority of the CMS and
eGateway disconnection alarm.
This setting is available only in
Continuous Monitoring
mode.
Alarm Delay
Alarm Delay
6 sec
•
1 sec
-
15 sec
for continuously
measured parameters, the monitor
does not present the alarm if the alarm
condition is resolved within the delay
time.
•
Off
: an alarm is always presented.
The setting of
Alarm Delay
is not applied
to the apnea alarms.
Other
SPO2 Desat
Alarm Off
Disable
Selects whether the SpO
2
Desat alarm
can be switched off.
•
Disable
: the SpO
2
Desat alarm cannot
be switched off.
•
Enable
: the SpO
2
Desat alarm can be
switched off.
Apnea Alarm
Off
Disable
Selects whether the apnea alarm can be
switched off.
•
Disable
: the apnea alarm cannot be
switched off.
•
Enable
: the apnea alarm can be
switched off.
Summary of Contents for VS 8A
Page 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 8: ...2 6 This page intentionally left blank ...
Page 26: ...1 6 This page intentionally left blank ...
Page 58: ...3 24 This page intentionally left blank ...
Page 66: ...5 4 This page intentionally left blank ...
Page 70: ...6 4 This page intentionally left blank ...
Page 98: ...8 16 This page intentionally left blank ...
Page 142: ...13 4 This page intentionally left blank ...
Page 162: ...15 16 This page intentionally left blank ...
Page 166: ...16 4 This page intentionally left blank ...
Page 202: ...19 8 This page intentionally left blank ...
Page 224: ...22 10 This page intentionally left blank ...
Page 250: ...B 8 This page intentionally left blank ...
Page 260: ...C 10 This page intentionally left blank ...
Page 271: ...D 11 This page intentionally left blank ...
Page 272: ...D 12 ...
Page 275: ...F 1 F Declaration of Conformity ...
Page 276: ...F 2 This page intentionally left blank ...
Page 277: ......
Page 278: ...P N 046 020097 00 4 0 ...