7 - 1
7
Alarms
7.1
Alarm Introduction
This chapter describes alarm functions and alarm settings. Physiological alarms are
available only when the monitor is in
Continuous Monitoring
mode.
7.2
Alarm Safety Information
WARNING
•
A potential hazard can exist if different alarm presets and default
configuration settings are used for the same or similar equipment in the
same care area, for example an intensive care unit or cardiac operating room.
•
If your monitor is connected to the central monitoring system (CMS), alarms
can be presented and controlled remotely. Remote suspension, inhibition, or
reset of monitor alarms via the CMS may cause a potential hazard. For more
information, see the operator’s manuals of the CMS.
•
The monitors in your care area may each have different alarm settings to suit
different patients. Always check that the alarm settings are appropriate for
your patient before start monitoring. Always make sure that necessary alarm
limits are active and set according to the patient's clinical condition.
•
Setting alarm limits to extreme values may cause the alarm system to
become ineffective. For example, high oxygen levels may predispose a
premature infant to retrolental fibroplasia. If this is a consideration do not
set the SpO
2
high alarm limit to 100%, which is equivalent to switching the
alarm off.
•
When the alarm sound is switched off, the monitor gives no alarm tones even
if a new alarm occurs. Be careful about whether to switch off the alarm sound
or not. When the alarms are off or while alarm audio is paused either
temporarily or indefinitely, observe the patient frequently.
•
When monitoring patients that are not continuously attended by a clinical
operator, properly configure the alarm system and adjust alarm settings as
per the patient's condition.
•
Do not rely exclusively on the audible alarm system for monitoring.
Adjustment of alarm volume to a low level may result in a hazard to the
patient. Always make sure that the audio alarm volume level is adequate in
your care environment. Always keep the patient under close surveillance.
Summary of Contents for VS 8A
Page 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 8: ...2 6 This page intentionally left blank ...
Page 26: ...1 6 This page intentionally left blank ...
Page 58: ...3 24 This page intentionally left blank ...
Page 66: ...5 4 This page intentionally left blank ...
Page 70: ...6 4 This page intentionally left blank ...
Page 98: ...8 16 This page intentionally left blank ...
Page 142: ...13 4 This page intentionally left blank ...
Page 162: ...15 16 This page intentionally left blank ...
Page 166: ...16 4 This page intentionally left blank ...
Page 202: ...19 8 This page intentionally left blank ...
Page 224: ...22 10 This page intentionally left blank ...
Page 250: ...B 8 This page intentionally left blank ...
Page 260: ...C 10 This page intentionally left blank ...
Page 271: ...D 11 This page intentionally left blank ...
Page 272: ...D 12 ...
Page 275: ...F 1 F Declaration of Conformity ...
Page 276: ...F 2 This page intentionally left blank ...
Page 277: ......
Page 278: ...P N 046 020097 00 4 0 ...