7 - 6
2.
Select the
Alarm
tab
→
Audio
tab.
3.
Set
High Alarm Interval
,
Med Alarm Interval
, and
Low Alarm Interval
:
7.6.2.3
Other Password Protected Audio Alarm Settings
The following alarm settings are password protected:
■
Minimum alarm volume
■
Alarm sound pattern
■
Alarm sound escalation switch and delay
7.6.3
Setting the Alarm Delay Time
For continuously measured parameters, you can set the alarm delay time. If the alarm
condition is resolved within the delay time, the monitor does not present the alarm.
This setting is password protected. For more information, see
.
The setting of
Alarm Delay
is not applied to the apnea alarms. You can set
Apnea Delay
separately.
WARNING
•
The alarm delay time can be set to 15 sec at most. Changing this setting to an
inappropriate level may result in a hazard to the patient.
7.6.4
Setting the Apnea Delay Time
To set the apnea delay time, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
Alarm
column select
Setup.
2.
Select
Apnea Delay
to set the apnea delay time.
7.6.5
Restoring the Default Alarm Settings
To reset all alarm settings to the defaults, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
Alarm
column select
Setup.
Item
Range
Default
High Alarm Interval
3 s to 15 s
10 s
Med Alarm Interval
3 s to 30 s
20 s
Med Alarm Interval
16 s to 30 s
20 s
Summary of Contents for VS 8A
Page 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 8: ...2 6 This page intentionally left blank ...
Page 26: ...1 6 This page intentionally left blank ...
Page 58: ...3 24 This page intentionally left blank ...
Page 66: ...5 4 This page intentionally left blank ...
Page 70: ...6 4 This page intentionally left blank ...
Page 98: ...8 16 This page intentionally left blank ...
Page 142: ...13 4 This page intentionally left blank ...
Page 162: ...15 16 This page intentionally left blank ...
Page 166: ...16 4 This page intentionally left blank ...
Page 202: ...19 8 This page intentionally left blank ...
Page 224: ...22 10 This page intentionally left blank ...
Page 250: ...B 8 This page intentionally left blank ...
Page 260: ...C 10 This page intentionally left blank ...
Page 271: ...D 11 This page intentionally left blank ...
Page 272: ...D 12 ...
Page 275: ...F 1 F Declaration of Conformity ...
Page 276: ...F 2 This page intentionally left blank ...
Page 277: ......
Page 278: ...P N 046 020097 00 4 0 ...