3 - 21
■
The alarm setup menus will not be displayed.
■
Alarm limits cannot be set.
Different features between continuous monitoring mode and spot check mode:
3.10.1.3
Standby Mode
You can temperately stop patient monitoring without switching off the monitor by
entering the standby mode. In standby mode, the monitor behaves as follows:
■
Stops all parameter measurement.
■
Disables all the alarms and prompt messages, except for the battery low alarm.
■
Turns screen brightness to the dimmest after entering the standby mode.
You can switch the monitor to standby mode manually, or set
Auto Enter Standby
for
the monitor to enter standby mode automatically.
To switch the monitor to standby mode:
1.
Press the power switch, or select the
Main Menu
quick key
→
from the
Patient
Management
column select
Standby
.
2.
If the monitor is in
Continuous Monitoring
mode, select a location in the drop
down list to define where the patient is when the monitor enters the standby
mode. If the monitor is in
Spot Check
mode, skip this step.
3.
Select
OK
.
To set
Auto Enter Standby
:
Functions
Continuous Monitoring mode
Spot check mode
Configure and use Sat-Seconds
(Nellcor)
Yes
No
Access
Alarm Setup
tab
Yes
No
Access
Events
tab
Yes
No
Access
Graphic Trends
tab
Yes
No
Auto data updating with the
central monitoring system
Yes
No
BP Averaging
No
Yes
Orthostatic BP measurement
No
Yes
Measure CO
2
(VS 9 series)
Yes
No
SpO
2
sourced RR (Mindray SpO
2
module)
No
Yes
Summary of Contents for VS 8A
Page 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 8: ...2 6 This page intentionally left blank ...
Page 26: ...1 6 This page intentionally left blank ...
Page 58: ...3 24 This page intentionally left blank ...
Page 66: ...5 4 This page intentionally left blank ...
Page 70: ...6 4 This page intentionally left blank ...
Page 98: ...8 16 This page intentionally left blank ...
Page 142: ...13 4 This page intentionally left blank ...
Page 162: ...15 16 This page intentionally left blank ...
Page 166: ...16 4 This page intentionally left blank ...
Page 202: ...19 8 This page intentionally left blank ...
Page 224: ...22 10 This page intentionally left blank ...
Page 250: ...B 8 This page intentionally left blank ...
Page 260: ...C 10 This page intentionally left blank ...
Page 271: ...D 11 This page intentionally left blank ...
Page 272: ...D 12 ...
Page 275: ...F 1 F Declaration of Conformity ...
Page 276: ...F 2 This page intentionally left blank ...
Page 277: ......
Page 278: ...P N 046 020097 00 4 0 ...