10 - 4
10.3.1.2
Taking a Temperature in Predictive Mode
1.
Verify that the probe is placed in the probe well.
2.
Verify that the temperature measurement type and site are correct.
3.
Unplug the probe from the probe well and the measurement mode changes to
Predictive
automatically.
4.
Insert it into a cover in the probe cover pack and press the probe handle down
firmly until the cover engages with the probe.
The temperature module starts to warm up when the probe is taken out. The
message “Temp Warming Up” displays in Temp parameter area. The warming up
time is about 2 seconds in room temperature. The monitor issues two beeps and
provides the message “Temp Prediction Ready” on the screen when warm-up is
complete. Then you can place the probe at measurement site.
5.
Place the probe at measurement site and wait till the measurement stabilizes.
When the dynamic symbol
appears, it indicates that the monitor starts to take
the measurement.
◆
When taking an oral temperature, apply the probe under the patient’s
tongue from either side of the mouth. Verify that the probe reaches the rear
sublingual pocket. Have the patient close his/her lips to hold the probe. Hold
the probe in place. Make sure that the probe contacts with the patient’s oral
tissue throughout the measurement.
◆
When taking an axillary temperature, lift the patient’s arm to expose the
entire armpit. Apply the probe as high as possible in the armpit. Check that
the probe tip is completely surrounded by the axillary tissue. Lower the
patient’s arm so that it is tightly placed at the patient side. Keep the patient’s
arm and the probe in place throughout the measurement.
◆
When taking a rectal temperature, separate patient's buttocks with one
hand, and the probe 1.5 cm inside the rectum with the other hand. For
pediatric patient, depth of insertion shall be less. Tilt the probe so that it
always contacts with patient’s tissue. Lubricant can be used in rectal mode.
The monitor will give a beep as the temperature measurement is complete. The
temperature reading displays continuously until the probe is taken out from the
probe well.
6.
Withdraw the probe. Press firmly the ejection button on the top of the probe to
eject the probe cover. Replace the probe into the probe well.
In
Predictive
mode, the monitor automatically enters
Monitor
mode in the following
cases:
■
Accurate temperature is not reached.
■
Neither measurement is taken nor is the probe replaced in the probe well in 60
seconds after the probe is withdrawn from the well.
The temperature type automatically changes to Predictive mode when the probe
returns to the probe well.
Summary of Contents for VS 8A
Page 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 8: ...2 6 This page intentionally left blank ...
Page 26: ...1 6 This page intentionally left blank ...
Page 58: ...3 24 This page intentionally left blank ...
Page 66: ...5 4 This page intentionally left blank ...
Page 70: ...6 4 This page intentionally left blank ...
Page 98: ...8 16 This page intentionally left blank ...
Page 142: ...13 4 This page intentionally left blank ...
Page 162: ...15 16 This page intentionally left blank ...
Page 166: ...16 4 This page intentionally left blank ...
Page 202: ...19 8 This page intentionally left blank ...
Page 224: ...22 10 This page intentionally left blank ...
Page 250: ...B 8 This page intentionally left blank ...
Page 260: ...C 10 This page intentionally left blank ...
Page 271: ...D 11 This page intentionally left blank ...
Page 272: ...D 12 ...
Page 275: ...F 1 F Declaration of Conformity ...
Page 276: ...F 2 This page intentionally left blank ...
Page 277: ......
Page 278: ...P N 046 020097 00 4 0 ...