11 - 15
5.
Evaluation result based on the set criteria:
◆
Negative
: the differences are within the acceptable range as per the set
criteria.
◆
Positive
: the differences are out of the acceptable range as per the set
criteria.
For details, see
11.10.4.5 Setting Orthostatic BP Evaluation Criterion
6.
Start/Stop standing blood pressure measurement before measurement and clear
data after measurement
7.
Start/Stop lying blood pressure measurement
8.
Add patient symptoms for result evaluation
11.10.3 Performing Orthostatic BP Measurement
Follow this procedure:
1.
Place the NIBP cuff as instructed in
.
2.
Swipe on the NIBP area to the left or right to switch to the orthostatic BP measure-
ment screen.
3.
Ask the patient to lie flat on the back and select
on the top to take lying BP
measurement. The monitor waits for a preset duration and starts measurement
automatically. The time duration is configurable. For details, see
.
4.
After the lying BP measurement is completed, ask the patient to stand up. Within 1
minute, select
on the bottom. The monitor automatically takes several times of
standing BP measurement, depending on your setting of
Standing BP Measure-
ment Interval
and
Maximum Standing BP Measurements
. For details, see
11.10.4.2 Setting Standing BP Measurement Interval
and
5.
Select symptoms if necessary. The list of symptoms are configurable. For details,
see
11.10.4.4 Setting Orthostatic BP Symptoms
.
6.
Upon completion of measurement, the evaluation result is displayed in the NIBP
area. Select
Negative
or
Positive
to check for details.
11.10.4 Changing Orthostatic BP Measurement Settings
11.10.4.1 Setting the Lying Duration of Patient
Follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
Configuration
column select
Workflow
→
input the required password
→
select
.
2.
Select
on the right of the workflow to be set. Make sure the workflow is
under
Spot Check
.
Summary of Contents for VS 8A
Page 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 8: ...2 6 This page intentionally left blank ...
Page 26: ...1 6 This page intentionally left blank ...
Page 58: ...3 24 This page intentionally left blank ...
Page 66: ...5 4 This page intentionally left blank ...
Page 70: ...6 4 This page intentionally left blank ...
Page 98: ...8 16 This page intentionally left blank ...
Page 142: ...13 4 This page intentionally left blank ...
Page 162: ...15 16 This page intentionally left blank ...
Page 166: ...16 4 This page intentionally left blank ...
Page 202: ...19 8 This page intentionally left blank ...
Page 224: ...22 10 This page intentionally left blank ...
Page 250: ...B 8 This page intentionally left blank ...
Page 260: ...C 10 This page intentionally left blank ...
Page 271: ...D 11 This page intentionally left blank ...
Page 272: ...D 12 ...
Page 275: ...F 1 F Declaration of Conformity ...
Page 276: ...F 2 This page intentionally left blank ...
Page 277: ......
Page 278: ...P N 046 020097 00 4 0 ...