12 - 2
12.3
Measurement Limitations
The following factors may influence the measurement accuracy:
■
Leaks or internal venting of sampled gas
■
Mechanical shock
■
Cyclic pressure up to 10 kPa (100 cmH
2
O)
■
Other sources of interference, if any
Measurement accuracy may be affected by the breath rate and inspiration/expiration
ratio (I: E ratio) as follow:
■
EtCO
2
value is within specification for breath rate ≤ 60 bpm and I/E ratio ≤ 1:1;
■
EtCO
2
value is within specification for breath rate ≤ 30 bpm and I/E ratio ≤ 2:1.
Measurement accuracy is unspecified for breath rate larger than 60 bpm.
12.4
CO
2
Display
The CO
2
numeric and waveform areas provide FiCO
2
measurement, EtCO
2
measurement, RR measurement and a CO
2
waveform.
1.
CO
2
waveform
2.
End tidal
CO
2
value (EtCO
2
)
3.
Fraction of inspired
CO
2
(FiCO
2
)
4.
RR (Respiration Rate)
5.
RR source: indicates the source of the current RR value
2
4
1
3
5
Summary of Contents for VS 8A
Page 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 8: ...2 6 This page intentionally left blank ...
Page 26: ...1 6 This page intentionally left blank ...
Page 58: ...3 24 This page intentionally left blank ...
Page 66: ...5 4 This page intentionally left blank ...
Page 70: ...6 4 This page intentionally left blank ...
Page 98: ...8 16 This page intentionally left blank ...
Page 142: ...13 4 This page intentionally left blank ...
Page 162: ...15 16 This page intentionally left blank ...
Page 166: ...16 4 This page intentionally left blank ...
Page 202: ...19 8 This page intentionally left blank ...
Page 224: ...22 10 This page intentionally left blank ...
Page 250: ...B 8 This page intentionally left blank ...
Page 260: ...C 10 This page intentionally left blank ...
Page 271: ...D 11 This page intentionally left blank ...
Page 272: ...D 12 ...
Page 275: ...F 1 F Declaration of Conformity ...
Page 276: ...F 2 This page intentionally left blank ...
Page 277: ......
Page 278: ...P N 046 020097 00 4 0 ...