Alarm Messages
This chapter lists only the physiological and technical alarm messages. Some messages
appearing on your monitor may not be included. In the “Cause and solution” column,
corresponding solutions are given instructing you to troubleshoot problems. If the prob-
lem persists, contact your service personnel.
D.1
Physiological Alarm Messages
D.1.1
General Physiological Alarm Messages
Note:
XX represents a measurement or parameter label, such as NIBP, RR, SpO
2
, CO
2
, PR, and so on.
D.1.2
SpO
2
Physiological Alarm Messages
D.1.3
PR Physiological Alarm Messages
Alarm
messages
Default
priority
Cause and solution
XX High
Med
XX value has risen above the high alarm limit or fallen
below the low alarm limit. Check the patient’s condition
and check if the patient category and alarm limit settings
are correct.
XX Low
Med
Alarm
message
Default
priority
Cause and solution
SpO2 Desat
High
The SpO
2
value falls below the desaturation alarm limit.
Check the patient’s condition and check if the alarm
limit settings are correct.
Alarm
message
Default
priority
Cause and solution
No Pulse
High
The pulse signal was so weak that the monitor cannot
perform pulse analysis. Check the patient’s condition,
SpO
2
sensor and measurement site.
Summary of Contents for VS 8A
Page 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 8: ...2 6 This page intentionally left blank ...
Page 26: ...1 6 This page intentionally left blank ...
Page 58: ...3 24 This page intentionally left blank ...
Page 66: ...5 4 This page intentionally left blank ...
Page 70: ...6 4 This page intentionally left blank ...
Page 98: ...8 16 This page intentionally left blank ...
Page 142: ...13 4 This page intentionally left blank ...
Page 162: ...15 16 This page intentionally left blank ...
Page 166: ...16 4 This page intentionally left blank ...
Page 202: ...19 8 This page intentionally left blank ...
Page 224: ...22 10 This page intentionally left blank ...
Page 250: ...B 8 This page intentionally left blank ...
Page 260: ...C 10 This page intentionally left blank ...
Page 271: ...D 11 This page intentionally left blank ...
Page 272: ...D 12 ...
Page 275: ...F 1 F Declaration of Conformity ...
Page 276: ...F 2 This page intentionally left blank ...
Page 277: ......
Page 278: ...P N 046 020097 00 4 0 ...