4 - 3
When a patient is discharged, all patient data, including patient information, trend data,
and physiological alarm information is deleted from the monitor. The technical alarms is
reset, and monitor settings returns to their defaults.
NOTE
•
Discharging a patient deletes all history data from the monitor.
4.3.1
Manually Discharging a Patient
Patients under
Continuous Monitoring
mode can be discharged manually. Follow this
procedure:
1.
Make sure all the patient data needed is properly saved.
2.
Select the patient information area at the top left corner of the screen.
3.
Select
Discharge Patient
.
4.3.2
Auto Discharging a Patient after Monitor Power Off
For patients under
Continuous Monitoring
mode, you can set the monitor to
automatically discharge a patient upon startup after being switched off for a period of
time. Follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required password
→
select
.
2.
Select the
Patient Management
tab
→
select the
Discharge
tab.
3.
Select a time for
Auto Discharge When Power Off
. The monitor will automatically
discharge the patient at next startup after being switched off for the selected
period of time. The default is
Never
. That is to say the monitor will not discharge a
patient upon startup no matter for how long the monitor has been switched off.
4.3.3
Changing Patient Management Settings
You can define which items can be displayed and edited from the
Patient Management
menu. To do so, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required password
→
select
.
2.
Select the
Patient Management
tab.
3.
Select the fields you want to display in the
Patient Management
menu.
4.
If necessary, select the customized fields and input names for these fields.
NOTE
•
If the monitor is connected with the CMS, the patient information items and
customized fields are loaded from the CMS.
Summary of Contents for VS 8A
Page 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 8: ...2 6 This page intentionally left blank ...
Page 26: ...1 6 This page intentionally left blank ...
Page 58: ...3 24 This page intentionally left blank ...
Page 66: ...5 4 This page intentionally left blank ...
Page 70: ...6 4 This page intentionally left blank ...
Page 98: ...8 16 This page intentionally left blank ...
Page 142: ...13 4 This page intentionally left blank ...
Page 162: ...15 16 This page intentionally left blank ...
Page 166: ...16 4 This page intentionally left blank ...
Page 202: ...19 8 This page intentionally left blank ...
Page 224: ...22 10 This page intentionally left blank ...
Page 250: ...B 8 This page intentionally left blank ...
Page 260: ...C 10 This page intentionally left blank ...
Page 271: ...D 11 This page intentionally left blank ...
Page 272: ...D 12 ...
Page 275: ...F 1 F Declaration of Conformity ...
Page 276: ...F 2 This page intentionally left blank ...
Page 277: ......
Page 278: ...P N 046 020097 00 4 0 ...