Installation
optoNCDT 1750
5.4.9
Connector and Sensor Cable
appr. 315
90 mm
3 ... 40 m
ILD1750 with pigtail
PC1700 with open ends
Never fall below the bending radius for the sensor cable of 55 mm (dynamic).
i
The fixed connected sensor cable is cable carriers suitable.
i
Unused open cable ends must be insulated to protect against short circuits or malfunction of the sen-
sor.
MICRO-EPSILON recommends to use the cable carriers suitable standard connection cable PC1700 of the
optional accessories, see Chap.
The connector and the cable component are marked with red markings which have to be aligned opposite
each other before connection. In addition, they come with guidance grooves to prevent them from being
wrongly connected. To release the plug-in connection, hold the plug-in connector on the grooved grips (outer
sleeves) and pull apart in a straight line. Pulling on the cable and the lock nut will only lock the plug-in con-
nector (ODU MINI-SNAP FP - lock) and will not release the connection.
Mount the cable connector if you use a cable carriers suitable sensor cable PC1700.
Avoid excessive pulling to the cables. Provide strain relieves near the connectors when cables > 5 m are
vertically free hanging.
Do not twist the connectors in opposite directions when connected.