Page 71
Setting Sensor Parameters
optoNCDT 1750
Measured
Value Counter
18 Bit
x = digital value
[0; 262143]
Time Stamp
2 words,
à 16 Bit
x = digital value Lo
y = digital value Hi
[0; 65535]
[0; 65535]
t
[µs]
=
( x)
1
1000
t = time stamp [µs]
[0; 1h11m34.967s]
Unlinearized
center of grav-
ity
18 Bit
x = digital value
[0; 262143]
CG
[%]
=
x
100
262143
CG = center of gravity [%]
[0; 100]
Video signal
16 Bit
512 pixels
[0; 1023]
Measurement
frequency
18 Bit
x = digital value
f = frequency [Hz]
[3000; 75000]
f
[Hz]
=
x
10
State information transferred in the distance value
Distance value
Description
262075
Too much data for selected baud rate
262076
There is no peak present
262077
Peak is located in front of the measuring range (MR)
262078
Peak is located after the measuring range (MR)
262080
Measurement value cannot be evaluated
262081
Peak is too wide
262082
Laser is off
Summary of Contents for optoNCDT 1750
Page 139: ......