Page 83
Setting Sensor Parameters
optoNCDT 1750
4 mA
3.8 mA
Reserve measuring range
0 %
16 %
60 %
50 %
MR
12 mA
20 mA
Analog
Out
20.2 mA
Standard output
Output characteristic af
ter zero setting
Output characteristic af
ter mastering
100 %
Out min
Out max
Master point Master value I
Out min
I
Out max
16 %
(8 mm)
0 mm
9.44 mA
(-8 mm)
20.0 mA
(25 mm)
60 %
(30 mm)
10 mm
4.00 mA
(-15 mm)
18.40 mA
(30 mm)
Fig. 43 Current output with zeroing or mastering
0 V
-0.1 V
Reserve measuring range
0 %
16 %
60 %
50 %
MR
5 V
10 V
Analog
Out
10.1 V
Standard output
Output characteristic af
ter zero setting
Output characteristic af
ter mastering
100 %
Out min
Out max
Master point Master value U
Out min
U
Out max
16 %
(8 mm)
0 mm
1.70 V
(-8 mm)
10.0 V
(25 mm)
60 %
(30 mm)
10 mm
0.00 V
(-15 mm)
9.00 V
(30 mm)
Fig. 44
Voltage output with zeroing or mastering;
U
OUT
= 0 ... 10 V
Summary of Contents for optoNCDT 1750