Seite 122
Anhang | ASCII Communication with Sensor
optoNCDT 1750
A 3.2.7
Key Function
A 3.2.7.1
KEYLOCK, Set Key lock
KEYLOCK NONE | ACTIVE | AUTO [<timeout period>]
Key lock configuration
- NONE: Key is active, no key lock
- ACTIVE: Key lock is activated immediately after restart
- AUTO: Key lock is only activated
<timeout period>
, 1 ... 60 minutes after reboot
A 3.2.8
Measurement
A 3.2.8.1
TARGETMODE, Measurement Task
TARGETMODE STANDARD | MULTISURFACE | PENETRATION
Choice of material dependent presets
- STANDARD: suitable for materials, e.g. made of ceramics, metal, plastics or wood
- MULTISURFACE: suitable for materials with changing surfaces, e.g. PCB or hybrid materials
- PENETRATION: suitable for materials with strong penetration depth of the laser light
A 3.2.8.2
MEASPEAK, Choice of the Peak in the Video Signal
MEASPEAK DISTA | DISTW | DIST1 | DISTL
- DISTA: output of peak with highest amplitude (standard)
- DISTW: output of peak with the largest area
- DIST1: output of first peak
- DISTL: output of last peak
A 3.2.8.3
MEASRATE, Measuring rate
MEASRATE <frequency>
Specifies the measuring rate in kHz, range 0.3 … 7.5 kHz.