Page 61
Setting Sensor Parameters
optoNCDT 1750
7.5
Signal Processing
7.5.1
Preliminary Remark
Change to the
Signal processing
menu in the
Settings
tab.
According to the previous setting in the
Diagram type
area, a diagram is displayed in the right part of the
display. The diagram is active and all settings become immediately visible. Notes on the chosen settings are
displayed below.
In the left area, the menus for the
signal processing
area are displayed.
7.5.2
Averaging
7.5.2.1
General
Averaging is recommended for static measurements or slowly changing measured values.
Averaging
Inactive
Measurement values are not averaged.
Moving N values
2 / 4 / 8 ... 128
Value
Indication of averaging mode. The averaging
number N indicates the number of consecu-
tive measurement values to be averaged in the
sensor.
Recursive N values
2 ... 32768
Value
Median N values
3 / 5 / 7 / 9
Value
Measurement averaging is performed after the distance values have been calculated, and before they are
issued through the relevant interfaces.
Averaging
- improves the resolution,
- allows masking individual interference points, and
- ‘smoothes’ the reading.
Linearity is not affected by averaging.
The average values are continuously recalculated with each measurement. The desired averaging depth
is only achieved after the number of recorded measurement values corresponds at least to the averaging
depth.
i
The defined type of average value and the averaging number must be stored in the sensor to ensure
they are hold after it is switched off.
Grey shaded
fields require a
selection.
Value
Dark-bordered
fields require
you to specifie
a value.