Page 48
Operation
optoNCDT 1750
1
Stop
stops the diagram; data selection and zoom function are still possible.
Save
opens the Windows
selection dialog for file name and storage location to save the video signal in a CSV file.
2
The video curves to be displayed while or after measurement can be switched on or off. Non active
curves are highlighted in gray and can be added by clicking the hook. If you only want to see a single
signal, then click on its name.
- Peak marking (vertical blue line), corresponds to the calculated measurement value
- Linearized measuring range (limited by means of gray shading), cannot be changed
- Masked range (limited by means of light blue shading), changeable
3
For scaling the measurement value axis (y-axis) of the graphics you can either choose
Auto
(= auto
scaling) or
Manual
(= manual setting).
4
The search function enables time-saving access to functions and parameters.
i
ASCII commands to the sensor can also be sent via the search function.
5
In the text boxes above the graphics current values for distance, exposure time and current measuring
rate are displayed.
6
Mouse over function. When moving the mouse over the graphic in stopped state curve points are
marked with a circle symbol and the related intensity is displayed. The related x position in % appears
above the graphic filed.
7
The linearized range is between the gray shading in the diagram and cannot be changed.
Only peaks
which centers are within this range can be calculated as measurement value. The masked range can be
limited on request and is additionally limited by means of a light blue shading on the right and on the
left. The remaining peaks in the resulting range are used for evaluation.
8
Scaling of the x-axis can be defined by means of a input field below the time axis.
9
Scaling of the x-axis: you can enlarge (zoom) the overall signal by means of the left slider during ongo-
ing measurement. If the diagram is stopped, you can also use the right slider. The zoom window can
also be moved by means of the mouse in the middle of the zoom window (arrow cross).
10 Choice of a diagram type: measurement or video signal representation.
By displaying the video signal, you can detect effect of the adjustable measurement task (target material),
choice of peak and possible interfering signals by means of reflections.