Page 54
Setting Sensor Parameters
optoNCDT 1750
7.4
Data Recording
7.4.1
Preliminary Remark
Change to the
Data recording
menu in the
Settings
tab .
According to the previous setting in the
Diagram type
area, a diagram is displayed in the right part of the
display. The diagram is active and all settings become immediately visible. Notes on the chosen settings are
displayed below.
In the left area, the menus for the
Data recording
are displayed.
7.4.2
Measuring Rate
The measuring rate indicates the number of measurements per second.
Select the required measuring rate.
Measuring
rate
300 Hz / 625 Hz / 1.25 kHz /
2.5 kHz / 5 kHz / 7.5 kHz /
Use a high measuring rate for bright and mat measurement ob-
jects. Use a low measuring rate for dark or shiny measurement
objects (e.g. black painted surfaces) to improve the measure-
ment result.
Free measuring
rate
Value
At a maximum measuring rate of 7.5 kHz, the CMOS element is exposed 7500 times per second. The lower
the measuring rate, the longer the maximum exposure time.
The measuring rate is factory set to 2.5 kHz.
Grey shaded
fields require a
selection.
Value
Dark-bordered
fields require
you to specifie
a value.