Page 79
Setting Sensor Parameters
optoNCDT 1750
Voltage output (with mastering), reference value midrange
Variables
Value range
Formula
U
OUT
= Voltage [V]
[-0.05; <0] SMR reserve
[0; 5] Measuring range
[>5; 5.05] EMR reserve
d
[mm]
=
* MR
[mm]
(U
OUT
[V]
- 2.5)
5
d
[mm]
=
* MR
[mm]
(U
OUT
[V]
- 5)
10
[-0.1; <0] SMR reserve
[0; 10] Measuring range
[>10; 10.1] EMR reserve
MR = Measuring range [mm]
{2/10/20/50/100/200}
MP = Master position [mm]
[0; MR]
d = Distance [mm]
for MP
£
0.5MR: [-MP; 0.5MR]
for MP > 0.5MB: [-0.5MR; MR - MP]
Voltage output (with teaching)
Variables
Value range
Formula
U
OUT
= Voltage [V]
[-0.05; <0] SMR reserve
[0; 5] Measuring range
[>5; 5.05] EMR reserve
d
[mm]
=
5
*|n
[mm]
- m
[mm]
|
U
OUT
[V]
d
[mm]
=
10
*|n
[mm]
- m
[mm]
|
U
OUT
[V]
[-0.1; <0] SMR reserve
[0; 10] Measuring range
[>10; 10.1] EMR reserve
MR = Measuring range [mm] {2/10/20/50/100/200}
m, n = Teaching range [mm]
[0; MR]
d
= Distance [mm]
[m; n]
Summary of Contents for optoNCDT 1750