Anhang | Control Menu
optoNCDT 1750
A 4.2.2
Data Recording
Measuring rate
300 Hz / 625 Hz / 1.25 kHz /
2.5 kHz / 5 kHz / 7.5 kHz /
Use a high measuring rate for bright and mat measurement objects. Use
a low measuring rate for dark or shiny measurement objects (e.g. black
painted surfaces) to improve the measurement result.
Free measuring rate
Value
Input trigger
Output trigger
Trigger source
Multi-function in-
put / Synchroniza-
tion input
Trigger type
Edge / Level
Trigger level
high rising edge / low falling edge
Number of mea-
sured values
Infinite
Manual selection
Value Range: 1 ... 16382
Software
Trigger type
Edge / Level
Trigger level
high rising edge / low falling edge
Number of mea-
sured values
Infinite
Manual selection
Value Range: 1 ... 16382
Start triggering
Button starts data recording
Stop triggering
Sensor outputs continuous data
Inactive
No triggering
Masked area
Start of range
0 ... 99 %
Value
Setting the evaluation range for the „Region of interest“, i.e. only this
range is used for logging the measuring values. The
Start of range
value has to be smaller than the
End of range
value.
End of range
1 ... 100 %
Value