A 3.2.1.8
SYNC
SYNC NONE | MASTER | MASTER_ALT | SLAVE | SLAVE_ALT | SLAVE_MFI
Setting the type of synchronization:
- NONE: No synchronization
- MASTER: The sensor is master, i.e. it transmits synchronization pulses on the output.
- MASTER_ALT: The sensor is master, i.e. it transmits synchronization pulses with every 2nd cycle. Both sen-
sors measure alternately, e.g. thickness measurement using 2 sensor on transparent material
- SLAVE: Sensor is slave and expects synchronization pulses from another optoNCDT 1750.
- SLAVE_ALT: Sensor is slave and expects synchronization pulses from a master sensor. Both sensors mea-
sure alternately, e.g. thickness measurement using 2 sensor on transparent material
- SLAVE_MFI: Sensor is slave and expects synchronization pulses from a master sensor at the multi-func-
tion input. Synchronization is performed with rising edge.