Installation and Start-Up | 3-31
To ensure safe operating conditions, the analyzer will not power up until
the MiniPurge
®
system has been connected to the instrument air sup-
ply (
ALARM/PRESSURIZED
indicator is GREEN) and has successfully
completed its purge timing cycle (the
PURGE TIMING/STATUS
LEDs
are now off) and the enclosure is pressurized.
After powering up the analyzer for the first time, certain checks must be
performed with the Electronic Enclosure door open to ensure the ana-
lyzer is operating properly. These checks are discussed in the following
power-up procedure. If problems are encountered during power-up, refer
to “Start-Up Diagnostic Checklist” in this chapter, for help in diagnosing
and correcting problems.
Always disconnect main AC power and/or external power sources
to the analyzer before opening any covers or doors on the analyzer to
check or perform maintenance on any components within the enclo-
sures.
If it is necessary to open the analyzer’s covers or doors while the
circuits are live, test the area for flammable gases (and proceed only
when the area is safe).
Purged Analyzer (Hazardous Location) Applications
To work on the analyzer with it powered up and its Electronics
Enclosure door open, the Purge Bypass Switch must be in the
“
BYPASS
” position.
When the Electronics Enclosure door is open, take appropriate precau-
tions to avoid electrical shock. Hazardous voltages are present inside.
Do not apply power to the analyzer if any of its flamepaths appear to
be scratched, dented, or worn. Applying power to an analyzer with a
damaged flamepath is dangerous and could result in serious injury
or death, and/or serious damage to equipment. See “Examining and
Caring for the Flamepaths” in Chapter 5.
Replace parts immediately if damage or wear is apparent. Contact
AMETEK if there is any doubt about the integrity of any flamepath.
!
WARNING
!
WARNING
Figure 3-10.
Purge Bypass Switch
label.
BY
PA
SS
ACTIV
E
Summary of Contents for 920
Page 39: ...Specifications 2 11 ...
Page 40: ...2 12 Model 920 Multi Gas Analyzer ...
Page 41: ...Specifications 2 13 2019 02 05 ...
Page 42: ...2 14 Model 920 Multi Gas Analyzer ...
Page 43: ...Specifications 2 15 ...
Page 44: ...2 16 Model 920 Multi Gas Analyzer ...
Page 45: ...Specifications 2 17 ...
Page 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Page 48: ...2 20 Model 920 Multi Gas Analyzer ...
Page 49: ...Specifications 2 21 ...
Page 50: ...2 22 Model 920 Multi Gas Analyzer ...
Page 51: ...Specifications 2 23 ...
Page 52: ...2 24 Model 920 Multi Gas Analyzer ...
Page 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Page 55: ...Specifications 2 27 ...
Page 56: ...2 28 Model 920 Multi Gas Analyzer ...
Page 57: ...Specifications 2 29 ...
Page 58: ...2 30 Model 920 Multi Gas Analyzer ...
Page 59: ...Specifications 2 31 ...
Page 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Page 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Page 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Page 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Page 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Page 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Page 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Page 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Page 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Page 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Page 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Page 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Page 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Page 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Page 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Page 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Page 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...