5-26 | Model 920 Multi-Gas Analyzer
PMT Level and PMT Balance
PMT Level
(
PmtLvl
) and
PMT Balance
(
PmtBal
) are adjusted only dur-
ing the
Auto-Setup
. The typical range of operation for both parameters
is 0–10 VDC. After a successful Auto-Setup, normal values for
PMT Level
should be 7.2–7.4 VDC while the
PMT Balance
should be between 4.0–7.0
VDC. The analyzer will function with values outside this range, but it is a
sign that a change has likely occurred in the transmission of light within
the Optical Bench (for example: dirty Windows, weak lamps, weak PMTs,
faulty
PMT Buffer
board – see “Troubleshooting and Diagnostics” in this
chapter for alarm conditions and corrective action.
At the end of every Auto-Setup, always check the PMT Level and PMT
Balance. Take note of severe changes in these parameters because they are
as important as the absolute value.
The
PMT Level
signal adjusts the gain of both PMTs simultaneously in the
same direction.
PMT Balance
adjusts the gain of the two PMTs in opposite
directions by controlling a virtual ground circuit on the
Optical Bench
board.
It is possible to achieve an acceptable completion number, yet have an
unacceptable PMT Level and/or PMT Balance signal. Always check both
of these values after an Auto-Setup.
If the PMT signals are outside their normal range of operation after the
Auto-Setup
is complete, check for the following:
• Adjust the
Lamp Max
setting slightly (typical range is 4.5–7.8 VDC)
and start another Auto-Setup.
• Was the Auto-Setup done with Zero gas flowing through the sample
system?
• Are there any PMT- or lamp-related alarms? If so, see
“Troubleshooting and Diagnostics” in this chapter for alarm corrective
action.
• Are the Measuring Cell Windows clean?
• Are the PMT signals weak?
• Do the source lamps need replacement (weak lamps)? If not, are the
source lamps aligned?
• If the lamps were replaced, is the correct socket connected to the cor-
rect lamp? Lamp 1 is the closest to the
Optical Bench
board.
(PmtLvl) RUN
F1 4
(PmtBal) RUN
F1 5
Summary of Contents for 920
Page 39: ...Specifications 2 11 ...
Page 40: ...2 12 Model 920 Multi Gas Analyzer ...
Page 41: ...Specifications 2 13 2019 02 05 ...
Page 42: ...2 14 Model 920 Multi Gas Analyzer ...
Page 43: ...Specifications 2 15 ...
Page 44: ...2 16 Model 920 Multi Gas Analyzer ...
Page 45: ...Specifications 2 17 ...
Page 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Page 48: ...2 20 Model 920 Multi Gas Analyzer ...
Page 49: ...Specifications 2 21 ...
Page 50: ...2 22 Model 920 Multi Gas Analyzer ...
Page 51: ...Specifications 2 23 ...
Page 52: ...2 24 Model 920 Multi Gas Analyzer ...
Page 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Page 55: ...Specifications 2 27 ...
Page 56: ...2 28 Model 920 Multi Gas Analyzer ...
Page 57: ...Specifications 2 29 ...
Page 58: ...2 30 Model 920 Multi Gas Analyzer ...
Page 59: ...Specifications 2 31 ...
Page 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Page 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Page 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Page 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Page 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Page 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Page 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Page 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Page 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Page 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Page 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Page 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Page 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Page 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Page 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Page 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Page 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...